| Literature DB >> 25460893 |
Satish K Tuteja1, Manil Kukkar2, C R Suri3, A K Paul2, Akash Deep4.
Abstract
2-Aminobenzyl amine (2-ABA) functionalized graphene is proposed for the ultrasensitive immunosensing of Cardiac Troponin I (cTnI). 2-ABA was electrochemically polymerized on the graphene decorated interdigitated electrode to obtain the amine functionalized graphene (f-GN). The f-GN electrode was then modified with monoclonal anti-cTnI antibodies via Schiff reaction based chemistry. Detailed characteristics of the processes involved and the finally developed antibody conjugated f-GN interdigitated electrode have been studied. The above micro-device was used in a drain source configuration for the sensing of cTnI. A wide dynamic linear range of antigen detection (0.01-1ng/mL) is achieved with the limit of detection of 0.01ng/mL. The utility of the proposed sensing technique is demonstrated by successfully testing the antigen concentration in spiked serum samples.Entities:
Keywords: 2-Aminobenzyl amine; Biomarker; Cardiac Troponin I; Graphene; Immunosensor
Mesh:
Substances:
Year: 2014 PMID: 25460893 DOI: 10.1016/j.bios.2014.10.072
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618