Literature DB >> 25456759

Multicenter evaluation of a next-generation balloon-expandable transcatheter aortic valve.

John Webb1, Gino Gerosa2, Thierry Lefèvre3, Jonathon Leipsic4, Mark Spence5, Martyn Thomas6, Matthias Thielmann7, Hendrik Treede8, Olaf Wendler9, Thomas Walther10.   

Abstract

BACKGROUND: The SAPIEN 3 (Edwards Lifesciences Inc., Irvine, California) transcatheter valve incorporates features designed to address the well-known deficiencies of transcatheter aortic valve replacement (TAVR). An ultra-low-profile delivery system facilitates safe, controlled, and accurate implantation and an external seal minimizes paravalvular regurgitation.
OBJECTIVES: The study evaluated whether TAVR with this third-generation valve would be a viable alternative to high- or intermediate-risk surgery for severe aortic stenosis.
METHODS: The prospective study enrolled 150 patients at 16 sites in Europe and Canada. Clinical and echocardiographic outcomes were assessed at baseline, post-procedure, and 30 days. New sizing recommendations were developed during the course of the study.
RESULTS: Patients were 83.6 ± 5.0 years of age, with multiple comorbidities reflected by a Society of Thoracic Surgeons score of 7.4 ± 4.5% and logistic EuroSCORE of 21.6 ± 12.3%. A transfemoral approach was chosen in 64.0% and alternative access (transapical/direct aortic) in the remainder. At 30 days, paravalvular regurgitation was none to mild in 96.4% and moderate in 3.5%. No patient had severe regurgitation. Transfemoral implantation was associated with low mortality (2.1%), no disabling stroke (0.0%), and fully percutaneous access and closure in 95.8%. Nontransfemoral alternative access was associated with higher rates of mortality (11.6%) and stroke (5.6%).
CONCLUSIONS: This third-generation device addresses major deficiencies of earlier valves in terms of ease of use, accuracy of positioning, and paravalvular sealing. The rates of mortality and stroke with transfemoral access are among the lowest reported and support further evaluation as an alternative to open surgery in intermediate-risk patients. (Safety and Performance Study of the Edwards SAPIEN 3 Transcatheter Heart Valve [SAPIEN3]; NCT01808287).
Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  aortic stenosis; transcatheter aortic valve replacement

Mesh:

Year:  2014        PMID: 25456759     DOI: 10.1016/j.jacc.2014.09.026

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  38 in total

Review 1.  Transcatheter Aortic Valve Replacement: Comprehensive Review and Present Status.

Authors:  Sameer Arora; Jacob A Misenheimer; Radhakrishnan Ramaraj
Journal:  Tex Heart Inst J       Date:  2017-02-01

2.  [We carry out too many TAVI - contra].

Authors:  S Yücel; H Ince; G D'Ancona; S Kische; A Öner
Journal:  Herz       Date:  2016-03       Impact factor: 1.443

3.  Comparison of outcomes using balloon-expandable versus self-expanding transcatheter prostheses according to the extent of aortic valve calcification.

Authors:  Won-Keun Kim; Johannes Blumenstein; Christoph Liebetrau; Andreas Rolf; Luise Gaede; Arnaud Van Linden; Mani Arsalan; Mirko Doss; Jan G P Tijssen; Christian W Hamm; Thomas Walther; Helge Möllmann
Journal:  Clin Res Cardiol       Date:  2017-08-09       Impact factor: 5.460

4.  Onsite cardiac surgery standby during transcatheter aortic valve implantation: when and why.

Authors:  Marco Vola; Antoine Gerbay
Journal:  J Thorac Dis       Date:  2016-10       Impact factor: 2.895

5.  Transcatheter aortic valve replacement: favorable clinical outcomes support role in intermediate risk surgical patients.

Authors:  Ravilla Mahidhar; Jon R Resar
Journal:  J Thorac Dis       Date:  2016-09       Impact factor: 2.895

Review 6.  [Interdisciplinary differential treatment of structural heart disease : When operation and when catheter-based intervention?].

Authors:  S Yücel; H Ince; S Kische; M A Sherif; H Bushnaq; A Bärisch; A Öner
Journal:  Herz       Date:  2016-08       Impact factor: 1.443

Review 7.  Evolving Indications for Transcatheter Aortic Valve Interventions.

Authors:  Anna Franzone; Thomas Pilgrim; Stefan Stortecky; Stephan Windecker
Journal:  Curr Cardiol Rep       Date:  2017-09-14       Impact factor: 2.931

8.  Permanent Pacemaker Implantation after TAVR - Predictors and Impact on Outcomes.

Authors:  Marcel Weber; Jan-Malte Sinning; Christoph Hammerstingl; Nikos Werner; Eberhard Grube; Georg Nickenig
Journal:  Interv Cardiol       Date:  2015-05

9.  Balloon- or Self-Expandable TAVI: Clinical Equipoise?

Authors:  John Jose; Gert Richardt; Mohamed Abdel-Wahab
Journal:  Interv Cardiol       Date:  2015-05

10.  The Application of Frailty to the Modern Cardiac Risk Assessment: a Case-Based Review.

Authors:  Matthew Finn; Philip Green
Journal:  Curr Cardiovasc Risk Rep       Date:  2015-10-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.