Literature DB >> 25456513

Prevention of airway fires: testing the safety of endotracheal tubes and surgical devices in a mechanical model.

Soham Roy1, Lee P Smith2.   

Abstract

OBJECTIVES: This study was designed to assess the ability of carbon dioxide (CO2) lasers and radiofrequency ablation devices (Coblator) (ArthoCare Corporation, Sunnyvale, CA) to ignite either a non-reinforced (polyvinylchloride) endotracheal tube (ETT) or an aluminum and fluoroplastic wrapped silicon ("laser safe") ETT at varying titrations of oxygen in a mechanical model of airway surgery.
METHODS: Non-reinforced and laser safe ETTs were suspended in a mechanical model imitating endoscopic airway surgery. A CO2 laser set at 5-30 watts was fired at the ETT at oxygen concentrations ranging from 21% to 88%. The process was repeated using a radiofrequency ablation (RFA) device. All trials were repeated to ensure accuracy.
RESULTS: The CO2 laser ignited a fire when contacting a non-reinforced ETT in under 2 seconds at oxygen concentrations as low as 44%. The CO2 laser could not ignite a laser safe ETT under any conditions, unless it struck the non-reinforced distal tip of the ETT. With the RFA, a fire could not be ignited with either reinforced or non-reinforced ETTs.
CONCLUSIONS: RFA presents no risk of ignition in simulated airway surgery. CO2 lasers should be utilized with a reinforced ETT or no ETT, as fires can easily ignite when lasers strike a non-reinforced ETT. Decreasing the fraction of inspired oxygen reduces the risk of fire.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25456513     DOI: 10.1016/j.amjoto.2014.10.009

Source DB:  PubMed          Journal:  Am J Otolaryngol        ISSN: 0196-0709            Impact factor:   1.808


  4 in total

1.  Risks and prevention of surgical fires : A systematic review.

Authors:  I Kezze; N Zoremba; R Rossaint; A Rieg; M Coburn; G Schälte
Journal:  Anaesthesist       Date:  2018-05-15       Impact factor: 1.041

2.  Fire safety study on high-flow nasal oxygen in shared-airway surgeries with diathermy and laser: simulation based on a physical model.

Authors:  Man-Yun Chang; Jui-Hung Chen; Shih-Pin Lin; Wei-Nung Teng; Shu-Wei Liao; Chien-Kun Ting; Mei-Yung Tsou; Hui-Hua Kenny Chiang; Fu-Wei Su
Journal:  J Clin Monit Comput       Date:  2021-03-30       Impact factor: 1.977

3.  Supraglottoplasty in neonates and infants: A radiofrequency ablation approach.

Authors:  Shilei Pu; Hongming Xu; Xiaoyan Li
Journal:  Medicine (Baltimore)       Date:  2018-02       Impact factor: 1.817

4.  Safe Use of Flammable Endotracheal Tubes During Anesthesia for Laryngeal Laser Surgery-Report of 1024 Cases and a Brief Literature Review.

Authors:  Masood Mohseni; Farzad Izadi
Journal:  Iran J Otorhinolaryngol       Date:  2019-07