Prospective, multi-center evaluation of a silicon carbide coated cobalt chromium bare metal stent for percutaneous coronary interventions: two-year results of the ENERGY Registry.

Raimund Erbel1, Holger Eggebrecht2, Ariel Roguin3, Erwin Schroeder4, Sebastian Philipp5, Thomas Heitzer6, Harald Schwacke7, Oded Ayzenberg8, Antonio Serra9, Nicolas Delarche10, Andreas Luchner11, Ton Slagboom12.   

Abstract

BACKGROUND: Novel bare metal stents with improved stent design may become a viable alternative to drug-eluting stents in certain patient groups, particularly, when long-term dual antiplatelet therapy should be avoided.
PURPOSE: The ENERGY registry aimed to assess the safety and benefits of a cobalt-chromium thin strut bare metal stent with a passive coating in a large series of patients under real-world conditions. METHODS AND MATERIALS: This prospective registry recruited 1016 patients with 1074 lesions in 48 centers from April to November 2010. The primary endpoint was the rate of major adverse cardiac events (MACEs), a composite of cardiac death, myocardial infarction and clinically driven target lesion revascularization.
RESULTS: More than half of the lesions (61.0%) were type A/B1 lesions, mean lesion length was 14.5±6.5mm and mean reference vessel diameter 3.2±0.5mm. MACE rates at 6, 12 and 24months were 4.9%, 8.1% and 9.4%, target lesion revascularization rates 2.8%, 4.9% and 5.4% and definite stent thrombosis rates 0.5%, 0.6% and 0.6%. Subgroups showed significant differences in baseline and procedural characteristics which did not translate into significantly different clinical outcomes. Specifically, MACE rates at 24months were 13.5% in diabetics, 8.6% in small stents and 9.6% in acute coronary syndrome patients.
CONCLUSION: The population of ENERGY reflects real-world conditions with bare metal stents being mainly used in simple lesions. In this setting, percutaneous coronary intervention using a cobalt-chromium thin strut bare metal stent with a passive coating showed very good results up to 24months. (ClinicalTrials.gov:NCT01056120) SUMMARY FOR ANNOTATED TABLE OF CONTENTS: The ENERGY international registry evaluated the safety and benefits of a cobalt-chromium thin strut bare metal stent with passive coating in 1016 patients under real-world conditions until 2years. Results were encouraging with a low composite rate of cardiac death, myocardial infarction and clinically driven target lesion revascularization, even in the pre-defined high risk groups of diabetes, stents ≤2.75mm and acute coronary syndrome.
Copyright © 2014 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bare metal stent; Passive coating; Silicon carbide

Mesh:

Substances:

Year:  2014        PMID: 25456414     DOI: 10.1016/j.carrev.2014.10.002

Source DB:  PubMed          Journal:  Cardiovasc Revasc Med        ISSN: 1878-0938


  1 in total

1.  Do ultrathin strut bare-metal stents with passive coating improve efficacy in large coronary arteries? Insights from the randomized, multicenter BASKET-PROVE trials.

Authors:  Kim Wadt Hansen; Raban Jeger; Rikke Sørensen; Christoph Kaiser; Matthias Pfisterer; Tor Biering-Sørensen; Louise Hougesen Bjerking; Søren Galatius
Journal:  BMC Cardiovasc Disord       Date:  2019-10-16       Impact factor: 2.298

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.