Measurement of in vitro and in vivo stent geometry and deformation by means of 3D imaging and stereo-photogrammetry.

Iwona Zwierzak1, Daria Cosentino, Andrew J Narracott, Philipp Bonhoeffer, Vanessa Diaz, John W Fenner, Silvia Schievano.   

Abstract

PURPOSE: To quantify variability of in vitro and in vivo measurement of 3D device geometry using 3D and biplanar imaging.
METHODS: Comparison of stent reconstruction is reported for in vitro coronary stent deployment (using micro-CT and optical stereo-photogrammetry) and in vivo pulmonary valve stent deformation (using 4DCT and biplanar fluoroscopy). Coronary stent strut length and inter-strut angle were compared in the fully deployed configuration. Local (inter-strut angle) and global (dog-boning ratio) measures of stent deformation were reported during stent deployment. Pulmonary valve stent geometry was assessed throughout the cardiac cycle by reconstruction of stent geometry and measurement of stent diameter.
RESULTS: Good agreement was obtained between methods for assessment of coronary stent geometry with maximum disagreement of +/- 0.03 mm (length) and +/- 3 degrees (angle). The stent underwent large, non-uniform, local deformations during balloon inflation, which did not always correlate with changes in stent diameter. Three-dimensional reconstruction of the pulmonary valve stent was feasible for all frames of the fluoroscopy and for 4DCT images, with good correlation between the diameters calculated from the two methods. The largest compression of the stent during the cardiac cycle was 6.98% measured from fluoroscopy and 7.92% from 4DCT, both in the most distal ring.
CONCLUSIONS: Quantitative assessment of stent geometry reconstructed from biplanar imaging methods in vitro and in vivo has shown good agreement with geometry reconstructed from 3D techniques. As a result of their short image acquisition time, biplanar methods may have significant advantages in the measurement of dynamic 3D stent deformation.

Entities:  

Mesh:

Year:  2014        PMID: 25450318     DOI: 10.5301/ijao.5000362

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  1 in total

1.  Novel Method to Track Soft Tissue Deformation by Micro-Computed Tomography: Application to the Mitral Valve.

Authors:  Eric L Pierce; Charles H Bloodworth; Ajay Naran; Thomas F Easley; Morten O Jensen; Ajit P Yoganathan
Journal:  Ann Biomed Eng       Date:  2015-11-09       Impact factor: 3.934

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.