| Literature DB >> 25445119 |
Daniel P Boyle1, Teresa R Zembower2, Chao Qi3.
Abstract
We evaluated the ability of the Vitek MS system to classify clinical pulmonary Mycobacterium avium complex isolates compared to multilocus sequence analysis. Vitek MS accurately identified 55% of the isolates as M. avium and 18% as M. intracellulare, but misidentified 24 (27%) Mycobacterium chimaera isolates as Mycobacterium intracellulare.Entities:
Keywords: Matrix-assisted laser desorption ionization-time of flight mass spectrometry; Multilocus sequencing; Mycobacterium avium complex; Mycobacterium chimaera; Real-time PCR
Mesh:
Substances:
Year: 2014 PMID: 25445119 DOI: 10.1016/j.diagmicrobio.2014.09.026