Accuracy of an electronic apex locator in the retreatment of teeth obturated with plastic or cross-linked gutta-percha carrier-based materials: an ex vivo study.

Manuele Mancini1, Pietro Palopoli2, Lorenzo Iorio2, Gabriele Conte2, Luigi Cianconi2.   

Abstract

INTRODUCTION: This ex vivo study evaluated the accuracy of the Root ZX electronic apex locator (EAL) (J Morita Corp, Kyoto, Japan) in determining the working length during retreatment of canals sealed with 2 different carrier-based obturating materials (ProTaper Obturator [Dentsply Maillefer, Ballaigues, Switzerland] and GuttaCore [Dentsply Maillefer]) and also evaluated whether they influenced its accuracy differently.
METHODS: Fifty extracted single-rooted human teeth with sound apices were selected for the study. We measured canals with the direct visual technique using a #10 K-file and shaped them with ProTaper Universal instruments (Dentsply Maillefer) up to this predetermined length. After instrumentation, we determined the canal length again using the visual technique (direct length [DL]). This value was considered the "reference point." Specimens were then placed in a tooth holder (Pro-Train; Simit Dental, Mantova, Italy), and the canal length was electronically measured (electronic length 1 [EL1]). Specimens were then obturated with ProTaper Obturators (group 1, n = 25) and GuttaCore (group 2, n = 25). Seven days later, a new electronic length was determined (electronic length 2 [EL2]) during retreatment procedures in the presence of the obturating material. EL1 and EL2 values were compared with DL using the Bland and Altman method. The different influences of the tested materials on the accuracy of the EAL were calculated with repeated measures analysis of variance.
RESULTS: Both EL1 and EL2 values provided a statistically significant overestimation of the actual canal length (DL [P < .05]), with the EL2 values higher from DL. The 2 different materials did not influence the accuracy of the EAL differently (P = .486).
CONCLUSIONS: The measurements obtained with the EAL tested during orthograde retreatments can lead clinicians to overinstrumentation and consequent overfilling of the endodontic space.
Copyright © 2014 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Electronic apex locator; GuttaCore; ProTaper Obturator; retreatment; working length

Mesh:

Substances:

Year:  2014        PMID: 25443281     DOI: 10.1016/j.joen.2014.07.035

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  1 in total

1.  The influence of two forms of chlorhexidine on the accuracy of contemporary electronic apex locators.

Authors:  Ewa Marek; Ryta Łagocka; Katarzyna Kot; Krzysztof Woźniak; Mariusz Lipski
Journal:  BMC Oral Health       Date:  2019-12-31       Impact factor: 2.757

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.