Literature DB >> 25434808

Evaluations of secondary cancer risk in spine radiotherapy using 3DCRT, IMRT, and VMAT: A phantom study.

Jalil ur Rehman1, Ramesh C Tailor2, Muhammad Isa3, Muhammad Afzal4, James Chow5, Geoffrey S Ibbott2.   

Abstract

This study evaluated the secondary cancer risk from volumetric-modulated arc therapy (VMAT) for spine radiotherapy compared with intensity-modulated radiotherapy (IMRT) and 3-dimensional conformal radiotherapy (3DCRT). Computed tomography images of an Radiological Physics Center spine anthropomorphic phantom were exported to a treatment planning system (Pinnacle(3), version 9.4). Radiation treatment plans for spine were prepared using VMAT (dual-arc), 7-field IMRT (beam angles: 110°, 130°, 150°, 180°, 210°, 230°, and 250°), and 4-field 3DCRT technique. The mean and maximum doses, dose-volume histograms, and volumes receiving more than 2 and 4Gy to organs at risk (OARs) were calculated and compared. The lifetime risk for secondary cancers was estimated according to the National Cancer Registry Programme Report 116. VMAT delivered the lowest maximum dose to the esophagus (4.03Gy), bone (8.11Gy), heart (2.11Gy), spinal cord (6.45Gy), and whole lung (5.66Gy) as compared with other techniques (IMRT and 3DCRT). The volumes of OAR (esophagus) receiving more than 4Gy were 0% for VMAT, 27.06% for IMRT, and up to 32.35% for 3DCRT. The estimated risk for secondary cancer in the respective OAR is considerably lower in VMAT compared with other techniques. The results of maximum doses and volumes of OARs suggest that the risk of secondary cancer induction for the spine in VMAT is lower than IMRT and 3DCRT, whereas VMAT has the best target coverage compared with the other techniques.
Copyright © 2015 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Dose-volume histogram; Organs at risk; VMAT

Mesh:

Year:  2014        PMID: 25434808     DOI: 10.1016/j.meddos.2014.10.003

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  3 in total

1.  Estimation of second cancer risk after radiotherapy for rectal cancer: comparison of 3D conformal radiotherapy and volumetric modulated arc therapy using different high dose fractionation schemes.

Authors:  Daniel R Zwahlen; Laura I Bischoff; Günther Gruber; Marcin Sumila; Uwe Schneider
Journal:  Radiat Oncol       Date:  2016-11-10       Impact factor: 3.481

2.  Radiation-Induced Second Cancer Risk from External Beam Photon Radiotherapy for Head and Neck Cancer: Impact on in-Field and Out-of-Field Organs

Authors:  Vasanthan Sakthivel; Ganesh Kadirampatti Mani; Sunil Mani; Raghavendiran Boopathy
Journal:  Asian Pac J Cancer Prev       Date:  2017-07-27

3.  Dosimetric, Radiobiological and Secondary Cancer Risk Evaluation in Head-and-Neck Three-dimensional Conformal Radiation Therapy, Intensity-Modulated Radiation Therapy, and Volumetric Modulated Arc Therapy: A Phantom Study.

Authors:  Jalil Ur Rehman; Muhammad Isa; Nisar Ahmad; Gulfam Nasar; H M Noor Ul Huda Khan Asghar; Zaheer Abbas Gilani; James C L Chow; Muhammad Afzal; Geoffrey S Ibbott
Journal:  J Med Phys       Date:  2018 Apr-Jun
  3 in total