Literature DB >> 25433279

Use of ultrasound debridement as an adjunctive tool for treating infected prosthetic vascular grafts in the lower extremities.

Michele Carmo1, Daniela Mazzaccaro1, Iacopo Barbetta1, Alberto M Settembrini2, Sergio Roveri1, Miriam Fumagalli1, Luca Tassinari1, Piergiorgio G Settembrini3.   

Abstract

BACKGROUND: To evaluate the role of an ultrasound (US) debridement system to treat conservatively patients with poor medical conditions who presented with infection of a prosthetic vascular graft in the lower extremities.
METHODS: Data of all patients who underwent debridement of the grafts and/or surrounding tissue using an ultrasonic generator (Genera, Italia Medica, Milan, Italy) were recorded and retrospectively reviewed. Based on cultures, patients received specific antibiotic therapy. Partial graft removal, sartorius muscle flap rotation, or negative pressure wound treatment (NPWT) was selectively used. Early and late morbidity and mortality and recurrence were analyzed.
RESULTS: Thirteen patients (median age, 72 years; range, 57-92 years; 8 men) were treated (12 Szilagyi grade III and 1 grade II infections) with US debridement without removing the graft (8 cases) or with partial excision and "in situ" reconstruction with a silver prosthetic graft (5 cases). Sartorius flap rotation was associated in 6 and NPWT in 1 case. One patient died perioperatively because of pulmonary edema because of sepsis secondary to treatment failure. Estimated freedom from reinfection was 90.9 ± 9% at 6 months and 77.9 ± 14% at 1 and 2 years. Estimated limb survival was 78.7 ± 13% at 6 months, 65.6 ± 16% at 1 year, and 52.5 ± 18% at 2 years.
CONCLUSIONS: US debridement proved to be a valuable aid in the treatment of patients with infected grafts and poor medical conditions. Used in conjunction with antibiotics, it allowed us to be more conservative without compromising the chance of success.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2014        PMID: 25433279     DOI: 10.1016/j.avsg.2014.11.001

Source DB:  PubMed          Journal:  Ann Vasc Surg        ISSN: 0890-5096            Impact factor:   1.466


  3 in total

1.  Treatment of prosthetic vascular graft infection in the groin with ultrasound debridement: A case report.

Authors:  Makoto Haga; Hidenori Inoue; Shunya Shindo
Journal:  Ann Med Surg (Lond)       Date:  2020-10-21

2.  The effect of 22.5 kHz low-frequency contact ultrasound debridement (LFCUD) on lower extremity wound healing for a vascular surgery population: A randomised controlled trial.

Authors:  Christine Anne Murphy; Pamela Houghton; Tim Brandys; Gregory Rose; Dianne Bryant
Journal:  Int Wound J       Date:  2018-01-15       Impact factor: 3.315

3.  Multiple Low Frequency Ultrasound Enhances Bactericidal Activity of Vancomycin against Methicillin-Resistant Staphylococcus aureus Biofilms.

Authors:  Jin Wang; Ke Wen; Xu Liu; Chun-Xiao Weng; Rui Wang; Yun Cai
Journal:  Biomed Res Int       Date:  2018-09-30       Impact factor: 3.411

  3 in total