Calibration and equivalency analysis of image plate scanners.

G Jackson Williams1, Brian R Maddox1, Hui Chen1, Sadaoki Kojima2, Matthew Millecchia3.   

Abstract

A universal procedure was developed to calibrate image plate scanners using radioisotope sources. Techniques to calibrate scanners and sources, as well as cross-calibrate scanner models, are described to convert image plate dosage into physical units. This allows for the direct comparison of quantitative data between any facility and scanner. An empirical relation was also derived to establish sensitivity response settings for arbitrary gain settings. In practice, these methods may be extended to any image plate scanning system.

Mesh:

Year:  2014        PMID: 25430350     DOI: 10.1063/1.4886390

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523