Perioperative versus postoperative measurement of Taylor Spatial Frame mounting parameters.

Sami Sökücü1, Bilal Demir2, Osman Lapçin2, Umut Yavuz2, Yavuz S Kabukçuoğlu2.   

Abstract

OBJECTIVE: The aim of this study was to determine the differences, if any, between application parameters for the Taylor Spatial Frame (TSF) system obtained during surgery under fluoroscopy and after surgery from digital radiography.
METHODS: This retrospective study included 17 extremities of 15 patients (8 male, 7 female; mean age: 21.9 years, range: 10 to 55 years) who underwent TSF after deformity and fracture. Application parameters measured by fluoroscopy at the end of surgery after mounting the fixator were compared with parameters obtained from anteroposterior and lateral digital radiographs taken 1 day after surgery.
RESULTS: Fixator was applied to the femur in 8 patients, tibia in 6 and radius in 3. Mean time to removal of the frame was 3.5 (range: 3 to 7) months. Mean perioperative anteroposterior, lateral and axial frame offsets of patients were 9.1 (range: 3 to 20) mm, 18.1 (range: 5 to 37) mm and 95.3 (range: 25 to 155) mm, respectively. Mean postoperative anteroposterior, lateral and axial frame offset radiographs were 11.8 (range: 2 to 30) mm, 18 (range: 6 to 47) mm and 109.5 (range: 28 to 195) mm, respectively. There was no statistically significant difference between the groups (p>0.05).
CONCLUSION: While measurements taken during operation may lengthen the duration in the operation room, fluoroscopy may provide better images and is easier to perform than digital radiography. On the other hand, there is no difference between measurements taken during perioperative fluoroscopy and postoperative digital radiography.

Entities:  

Mesh:

Year:  2014        PMID: 25429572     DOI: 10.3944/AOTT.2014.13.0080

Source DB:  PubMed          Journal:  Acta Orthop Traumatol Turc        ISSN: 1017-995X            Impact factor:   1.511


  3 in total

1.  The silhouette technique: improving post-operative radiographs for planning of correction with a hexapod external fixator.

Authors:  Jonathan Wright; Shiraz A Sabah; Shelain Patel; Gavin Spence
Journal:  Strategies Trauma Limb Reconstr       Date:  2017-05-12

2.  Improving postoperative radiographs for the parameter measurement of hexapod external fixator using an additional foot ring.

Authors:  Yanshi Liu; Kai Liu; Feiyu Cai; Tao Zhang; Aihemaitijiang Yusufu
Journal:  J Orthop Surg Res       Date:  2021-11-13       Impact factor: 2.359

3.  Treatment of open fractures with a computer-assisted external fixator system without the use of fluoroscopy.

Authors:  Adnan Kara; Haluk Celik; Ali Seker; Ozgur Karakoyun; Raffi Armagan; Ersin Kuyucu; Mehmet Erdil
Journal:  J Orthop Surg Res       Date:  2016-04-26       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.