Comparative analysis of visual field and image distortion in 30° and 70° arthroscopes.

Aashay L Kekatpure1, Arnold Adikrishna1, Ji-Ho Sun1, Gyeong-Bo Sim1, Jae-Myung Chun1, In-Ho Jeon2.   

Abstract

PURPOSE: There have been very few attempts to compare the visual fields and image distortion in arthroscopes. To better understand the images generated using existing arthroscopes, we performed image-mapping experiments to assess field of view and image distortion. The purpose of this study was to quantify and compare the visual fields obtained using 30° and 70° arthroscopes and assess image distortion in each arthroscope.
METHODS: A complete arthroscopy system was used in this study. To perform this quantitative analysis, we created a customized measurement device that consisted of three parts: (1) distance marker, (2) chessboard pattern, and (3) angle marker. Three observers collectively assessed the appropriate position of the arthroscope during simulation. For each scope, ten simulations were performed at distances between 1 and 3 cm. Using the obtained arthroscope images, field of view and image distortion were measured and calculated.
RESULTS: The field of view of the 70° arthroscope was 5, 10, and 15 mm wider in diameter in comparison with the 30° arthroscope at 1, 2, and 3 cm, respectively. Moreover, the 70° arthroscope had less 0.66, 0.13, and 0.26 pixels of root-mean-square distance than the 30° arthroscope at 1, 2, and 3 cm, respectively. The 70° arthroscope also contained 0.78 pixels less at the maximal error than the average 30° arthroscope. Therefore, the 70° arthroscope demonstrated less distortion than the 30° arthroscope. There was no significant difference between the two scopes with respect to median curvature measurement at 1-cm distance.
CONCLUSION: The 70° arthroscope demonstrates technical advantages over the 30° arthroscope, including a wider field of view and a less image distortion at the periphery. A wide angle and less image distortion can help better orient the surgeon within the joint cavity when a panoramic picture is needed to repair rotator cuff tears, in case of hip arthroscopy, or while treating the lesions of posterior horn of medial meniscus.

Keywords:  Arthroscope; Comparative analysis; Field of view; Image distortion

Mesh:

Year:  2014        PMID: 25423874     DOI: 10.1007/s00167-014-3452-0

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  11 in total

1.  History of arthroscopy.

Authors:  Dusan Marić; Gerardo Garces; Armando Martinez; Lazar Petković
Journal:  Med Pregl       Date:  2002 Jan-Feb

2.  Eugen Bircher (1882-1956) the first knee surgeon to use diagnostic arthroscopy.

Authors:  Christopher W Kieser; Robert W Jackson
Journal:  Arthroscopy       Date:  2003-09       Impact factor: 4.772

3.  Arthroscopic surgery.

Authors:  R A Marquez
Journal:  Surg Laparosc Endosc       Date:  1992-03

Review 4.  Instrumentation: endoscopes and equipment.

Authors:  Michael R Gaab
Journal:  World Neurosurg       Date:  2012-02-10       Impact factor: 2.104

5.  Use of the 70° arthroscope for improved visualization with common arthroscopic procedures.

Authors:  Asheesh Bedi; Joshua Dines; David M Dines; Bryan T Kelly; Stephen J O'Brien; David W Altchek; Answorth A Allen
Journal:  Arthroscopy       Date:  2010-09-19       Impact factor: 4.772

6.  Does extended-field-of-view sonography improve interrater reliability for the detection of rotator cuff muscle atrophy?

Authors:  Eoin C Kavanagh; George Koulouris; Laurence Parker; William B Morrison; Diane Bergin; Adam C Zoga; John A Dlugosz; Levon N Nazarian
Journal:  AJR Am J Roentgenol       Date:  2008-01       Impact factor: 3.959

7.  Giants of orthopaedic surgery: Masaki Watanabe MD.

Authors:  Marlene DeMaio
Journal:  Clin Orthop Relat Res       Date:  2013-05-24       Impact factor: 4.176

8.  High-precision camera distortion measurements with a "calibration harp".

Authors:  Zhongwei Tang; Rafael Grompone von Gioi; Pascal Monasse; Jean-Michel Morel
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2012-10-01       Impact factor: 2.129

9.  Arthroscopic visual field mapping at the periphery of the medial meniscus: a comparison of different portal approaches.

Authors:  B S Tolin; A A Sapega
Journal:  Arthroscopy       Date:  1993       Impact factor: 4.772

Review 10.  Arthroscopic visualization of subscapularis tendon lesions.

Authors:  Tal S David; Hugo Bravo; Razvan Scobercea
Journal:  Orthopedics       Date:  2009-09       Impact factor: 1.390

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.