Literature DB >> 25421639

Mechanical evaluation of hip cement spacer reinforcement with stainless steel Kirschner wires, titanium and carbon rods, and stainless steel mesh.

Nobuhiro Kaku1, Tomonori Tabata, Hiroshi Tsumura.   

Abstract

INTRODUCTION: In two-stage treatments for infections after total hip arthroplasty, antibiotic-loaded cement spacers help treat the infection by antibiotic elution and prevent contraction. However, such spacers are weak and may fracture while awaiting replacement, impairing functionality. We evaluated whether a Kirschner wire (K-wire) mounted into the spacer reinforced its strength along with the effects of the reinforcing material, position, and intensity.
METHODS: Spacers without reinforcing materials constituted the control group. As reinforcing materials, stainless steel K-wires (diameters 3 and 6 mm), titanium alloy and carbon fibers (diameter 3.175 mm), and stainless steel meshes (inner and outer diameters, 6 and 9 mm, respectively) were inserted into the spacer mold before filling with cement. The spacers complied with ISO 7206-4; a compressive load was applied using a testing machine with a velocity of 25.4 mm/min, and the maximum load was recorded. We used 1-3 K-wires positioned on the medial side, lateral side, neck only, and stem only and tested 3 specimens for each condition.
RESULTS: The control group withstood the highest load. Stainless steel was the strongest material; 3-mm K-wires in the neck and lateral side withstood a higher load. The computed tomography (CT) imaging revealed a cavity between the K-wires and cement. When K-wires were inserted along the whole length, despite cement fractures, continuity was maintained because of the reinforcing materials.
CONCLUSION: It is difficult to improve the reinforcing strength of spacers using K-wires; however, K-wires prevented dislocation of cement spacer fragments, which can help prevent contraction and facilitate spacer removal during replacement.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25421639     DOI: 10.1007/s00590-014-1567-0

Source DB:  PubMed          Journal:  Eur J Orthop Surg Traumatol        ISSN: 1633-8065


  28 in total

1.  Clinical effectiveness of antibiotic-impregnated cement spacers for the treatment of infected implants of the hip joint.

Authors:  Kengo Yamamoto; Naoki Miyagawa; Toshinori Masaoka; Yoichi Katori; Takaaki Shishido; Atsuhiro Imakiire
Journal:  J Orthop Sci       Date:  2003       Impact factor: 1.601

2.  Individual bone cement spacers (IBCS) for septic hip revision-preliminary report.

Authors:  C Schoellner; S Fuerderer; J-D Rompe; A Eckardt
Journal:  Arch Orthop Trauma Surg       Date:  2003-05-01       Impact factor: 3.067

3.  Mechanical evaluation of unipolar hip spacer constructs.

Authors:  Frederick J Kummer; Eric Strauss; Kevin Wright; Erik N Kubiak; Paul E Di Cesare
Journal:  Am J Orthop (Belle Mead NJ)       Date:  2008-10

4.  A technique for the fabrication of a reinforced moulded articulating cement spacer in two-stage revision total hip arthroplasty.

Authors:  Matthew Kent; Rajesh Rachha; Manoj Sood
Journal:  Int Orthop       Date:  2009-08-20       Impact factor: 3.075

5.  A modified cement spacer technique for infected total hip arthroplasties with significant bone loss.

Authors:  Oren Ben-Lulu; Ali Farno; Allan E Gross; David J Backstein; Yona Kosashvili; Oleg A Safir
Journal:  J Arthroplasty       Date:  2011-08-31       Impact factor: 4.757

6.  A cement spacer for two-stage revision of infected implants of the hip joint.

Authors:  M Leunig; E Chosa; M Speck; R Ganz
Journal:  Int Orthop       Date:  1998       Impact factor: 3.075

7.  Antibiotics-impregnated cement spacers in the first step of two-stage revision for infected totally replaced hip joints: report of ten trial cases.

Authors:  Katsuyuki Dairaku; Michiaki Takagi; Hiroyuki Kawaji; Kan Sasaki; Masaji Ishii; Toshihiko Ogino
Journal:  J Orthop Sci       Date:  2009-12-08       Impact factor: 1.601

8.  Use of a cement-loaded Kuntscher nail in first-stage revision hip arthroplasty for massive femoral bone loss secondary to infection: a report of four cases.

Authors:  Corey Richards; Chris J Bell; Sameer Viswanathan; Hugh English; Ross W Crawford
Journal:  J Orthop Surg (Hong Kong)       Date:  2010-04       Impact factor: 1.118

9.  Preformed gentamicin spacers in two-stage revision hip arthroplasty: functional results and complications.

Authors:  Christophe Pattyn; Thomas De Geest; Pieter Ackerman; Emmanuel Audenaert
Journal:  Int Orthop       Date:  2010-11-30       Impact factor: 3.075

10.  Complications after spacer implantation in the treatment of hip joint infections.

Authors:  Jochen Jung; Nora Verena Schmid; Jens Kelm; Eduard Schmitt; Konstantinos Anagnostakos
Journal:  Int J Med Sci       Date:  2009-09-02       Impact factor: 3.738