Comparison of total haemoglobin mass measured with the optimized carbon monoxide rebreathing method across different Radiometer™ ABL-80 and OSM-3 hemoximeters.

G Turner, A J Richardson, N S Maxwell, J S M Pringle.   

Abstract

A new Radiometer™ hemoximeter (ABL-80) has recently become available to measure carboxyhaemoglobin concentration for the optimized CO-rebreathing method (oCOR-method). Within the English Institute of Sport (EIS), hemoximeters are used in three different laboratories; therefore, precision and agreement of total haemoglobin mass (tHbmass) determination across sites is essential, and comparison to the previous model OSM-3 is desirable. Six male and one female (age 30 ± 6 years, body mass 78.1 ± 10.6 kg) undertook the oCOR-method. Venous blood (~2 ml) was sampled immediately before and at 7 min during the oCOR-method; with seven replicates from each time point simultaneously analysed on five different Radiometer™ hemoximeters [OSM-3(1), OSM-3(2), ABL-80(1), ABL-80(2) and ABL-80(3)]. There were no differences (p > 0.05) between Δ%HbCO or mean tHbmass analysed with five different hemoximeters (OSM-3(1): 886 ± 167 g; OSM-3(2): 896 ± 160 g: ABL-80(1): 904 ± 157 g; ABL-80(2): 906 ± 163 g: ABL-80(3): 906 ± 162 g). However, the Bland-Altman plot revealed that there was closer agreement between ABL-80 machines for tHbmass than for the OSM-3. The variance (i.e. % error) across replicate samples decreased as the number of samples increased, with the error derived from the 'worse-case' scenario (single samples) being 1.2 to 1.6 fold greater in the OSM-3 than the ABL-80. Although there were no differences in the average tHbmass measured with five different hemoximeters, the new ABL-80 were in better agreement with each other compared to the old OSM-3. Previously, five replicates were required to achieve a low error using the OSM-3; however, three replicates are sufficient with the ABL-80 model to produce an error of ≤ 1% in tHbmass.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25420054     DOI: 10.1088/0967-3334/35/12/N41

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  4 in total

1.  Hemolysis during and after 21 days of head-down-tilt bed rest.

Authors:  Guy Trudel; Hans K Uhthoff; Odette Laneuville
Journal:  Physiol Rep       Date:  2017-12

2.  Hemoglobin concentration, total hemoglobin mass and plasma volume in patients: implications for anemia.

Authors:  James M Otto; James O M Plumb; Eleri Clissold; Shriya B Kumar; Denis J Wakeham; Walter Schmidt; Michael P W Grocott; Toby Richards; Hugh E Montgomery
Journal:  Haematologica       Date:  2017-06-08       Impact factor: 9.941

3.  Replicating measurements of total hemoglobin mass (tHb-mass) within a single day: precision of measurement; feasibility and safety of using oxygen to expedite carbon monoxide clearance.

Authors:  James O M Plumb; Shriya Kumar; James Otto; Walter Schmidt; Toby Richards; Hugh E Montgomery; Mike P W Grocott
Journal:  Physiol Rep       Date:  2018-09

4.  Application of the optimized carbon monoxide rebreathing method for the measurement of total haemoglobin mass in chronic liver disease.

Authors:  James O M Plumb; James M Otto; Shriya B Kumar; Mark Wright; Walter Schmidt; Michael P W Grocott; Hugh E Montgomery
Journal:  Physiol Rep       Date:  2020-03
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.