| Literature DB >> 2541950 |
Abstract
We evaluated a new, simple, enzymatic kinetic method from Wako Chemicals GmbH in comparison with a kinetic Jaffé method by using isotope dilution-mass spectrometry (ID-MS) as a reference method. An ID-MS-calibrated serum standard was used. Both the enzymatic and the Jaffé method correlated well with ID-MS, except for sera with high concentrations of bilirubin. Ethyl acetoacetate, acetone, and glucose in serum interfered somewhat with the Jaffé method but not with the enzymatic method. We conclude that the present enzymatic method has merit as compared with a Jaffé method for routine work, but is more expensive.Entities:
Mesh:
Substances:
Year: 1989 PMID: 2541950
Source DB: PubMed Journal: Clin Chem ISSN: 0009-9147 Impact factor: 8.327