A slippery coaxial, 3-F catheter device was developed for selective angiography, embolization, and infusion of chemotherapeutic agents. The device comprises three parts: an 0.018-inch, plastic-coated guide wire; a 3-F coaxial catheter; and a 6.5-F guiding catheter. Both the guide wire and coaxial catheter are coated with hydrophilic polymer, which becomes slippery when immersed in water. Experience with four patients indicates the new coaxial catheter device is capable of traversing tortuous vessels smoothly and safely, and it permits infusion of chemotherapeutic agents and 1-mm gelatin particles.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2541465     DOI: 10.1148/radiology.171.3.2541465

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  2 in total

1.  New plastic-coated coaxial catheter system for superselective hepatic catheterization: technical note.

Authors:  K Takayasu; K Aoki; T Ichikawa; T Ohmura; R Sekiguchi; H Kasugai; T Terauchi
Journal:  Cardiovasc Intervent Radiol       Date:  1990-12       Impact factor: 2.740

2.  Segmental embolotherapy for hepatic cancer: keys to success.

Authors:  H Uchida; N Matsuo; H Sakaguchi; N Nagano; K Nishimine; H Ohishi
Journal:  Cardiovasc Intervent Radiol       Date:  1993 Mar-Apr       Impact factor: 2.740