Sensitivity enhancement in off-axis integrated cavity output spectroscopy.

R Centeno, J Mandon, S M Cristescu, F J M Harren.   

Abstract

We report on a detailed model of an improved three mirror off-axis integrated cavity output spectroscopy (OA-ICOS) setup, which re-injects the light reflected by the optical cavity. The model simulates the impact of design parameters on instrument sensitivity and can be used for any off-axis configuration. We demonstrate the application of this model for the real-time detection of ethylene with a pulsed quantum cascade laser (QCL). The three mirror OA-ICOS scheme provides a 10 times increase in signal-to-noise ratio as compared to standard OA-ICOS, resulting in a noise equivalent absorption sensitivity of 1.5 x 10(-8) cm(-1) Hz(-1/2).

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25402039     DOI: 10.1364/OE.22.027985

Source DB:  PubMed          Journal:  Opt Express        ISSN: 1094-4087            Impact factor:   3.894


  3 in total

1.  Sensitive Spectroscopy of Acetone Using a Widely Tunable External-Cavity Quantum Cascade Laser.

Authors:  Faisal Nadeem; Julien Mandon; Amir Khodabakhsh; Simona M Cristescu; Frans J M Harren
Journal:  Sensors (Basel)       Date:  2018-06-27       Impact factor: 3.576

2.  RES-Q-Trace: A Mobile CEAS-Based Demonstrator for Multi-Component Trace Gas Detection in the MIR.

Authors:  Norbert Lang; Uwe Macherius; Henrik Zimmermann; Sven Glitsch; Mathias Wiese; Jürgen Röpcke; Jean-Pierre H van Helden
Journal:  Sensors (Basel)       Date:  2018-06-27       Impact factor: 3.576

3.  Long-Term Stable Online Acetylene Detection by a CEAS System with Suppression of Cavity Length Drift.

Authors:  Qixin He; Qibo Feng; Jiakun Li
Journal:  Sensors (Basel)       Date:  2019-01-26       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.