Literature DB >> 2540127

Routine determination of radiochemical purity of 99mTc-MIBI.

A Proulx1, J R Ballinger, K Y Gulenchyn.   

Abstract

The manufacturer recommends that the radiochemical purity of 99mTc-MIBI be determined with alumina TLC plates developed in ethanol, but for routine use this is slow and requires care. After comparing a variety of systems, we obtained equivalent results using ITLC-SG strips developed with saline to determine [99mTc]pertechnetate and with acetone to determine 99mTc colloid; % 99mTc-MIBI was determined by difference from 100%. These separations are the opposite of those obtained with the routine polar 99mTc radiopharmaceuticals. Equivalent results were obtained even more rapidly by extracting 99mTc-MIBI from water into chloroform.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2540127     DOI: 10.1016/0883-2889(89)90182-2

Source DB:  PubMed          Journal:  Int J Rad Appl Instrum A        ISSN: 0883-2889


  2 in total

1.  Alternative chromatographic system for the quality control of lipophilic technetium-99m radiopharmaceuticals such as [(99m)Tc(MIBI)₆].

Authors:  D P Faria; C A Buchpiguel; F L N Marques
Journal:  Braz J Med Biol Res       Date:  2015-03-03       Impact factor: 2.590

2.  Labeling of tannic Acid with technetium-99m for diagnosis of stomach ulcer.

Authors:  I T Ibrahim; M El-Tawoosy; H M Talaat
Journal:  ISRN Pharm       Date:  2011-07-19
  2 in total