| Literature DB >> 25395266 |
Eric F Bernstein1, Jason D Bloom, Lisa D Basilavecchio, Jessica M Plugis.
Abstract
BACKGROUND AND OBJECTIVES: A sharply contoured cryolipolysis vacuum applicator was developed to improve fit and tissue draw in the abdomen and flanks to better accommodate a range of body types and a variety of treatment sites. This study was carried out to evaluate the safety and efficacy of the new applicator for treatment of flank fat ("love handles"). STUDY DESIGN/Entities:
Keywords: cryolipolysis; flanks; multiple cycles; non-invasive body contouring; non-surgical fat reduction
Mesh:
Year: 2014 PMID: 25395266 PMCID: PMC4413829 DOI: 10.1002/lsm.22302
Source DB: PubMed Journal: Lasers Surg Med ISSN: 0196-8092 Impact factor: 4.025
Subject Weight and BMI Data
| Treatment visit | 12-week follow-up visit | |||
|---|---|---|---|---|
| Subject | Weight (lb) | BMI | Weight (lb) | Weight change (lb) |
| 01 | 158.0 | 24.7 | 154.2 | −3.8 |
| 02 | 129.8 | 22.3 | 128.9 | −0.9 |
| 03 | 161.5 | 27.7 | 161.2 | −0.3 |
| 04 | 148.8 | 28.1 | 147.7 | −1.1 |
| 05 | 132.4 | 22.7 | 130.5 | −1.9 |
| 06 | 119.3 | 23.3 | 115.3 | −4.0 |
| 08 | 143.7 | 26.3 | 142.1 | −1.6 |
| 09 | 127.9 | 21.3 | 130.5 | 2.6 |
| 10 | 126.6 | 21.1 | 124.5 | −2.1 |
At the treatment visit, all subjects had BMI <29 and agreed to maintain weight over the course of the study by avoiding diet and exercise changes. At 12-week follow-up, all subjects maintained weight within 5 lbs. of baseline.
Independent Photographic Review and Improvement Score Data
| Independent photo review (correct/incorrect) | Improvement score (0–10) | ||||||
|---|---|---|---|---|---|---|---|
| Subject | Reviewer 1 | Reviewer 2 | Reviewer 3 | Reviewer 4 | Reviewer 2 | Reviewer 3 | Reviewer 4 |
| 01 | C | C | C | C | 6 | 5 | 3 |
| 02 | C | C | C | C | 3 | 4 | 5 |
| 03 | C | C | C | C | 4 | 2 | 2 |
| 04 | C | C | C | I | 3 | 4 | −2 |
| 05 | C | I | C | C | −3 | 5 | 5 |
| 06 | C | C | C | C | 7 | 8 | 6 |
| 08 | C | C | C | C | 7 | 5 | 6 |
| 09 | C | C | C | C | 6 | 5 | 5 |
| 10 | C | C | C | C | 5 | 4 | 5 |
Four blinded, independent physicians correctly identified 94% of baseline images; 34 of 36 correct. Three physicians assigned improvement scores from 0 (none) to 10 (complete), resulting in a mean improvement score of 4.3 + 1.4 (mean + sem).
Figure 1This 41 year-old female received two sequential cryolipolysis cycles to each flank. Clinical photographs show flank fat reduction between baseline (a, c) and 12 weeks post-treatment (b, d). This subject had a weight gain of only 2.6 lbs. from her initial visit until the end of the study.
Figure 2This 42 year-old female received two sequential cryolipolysis cycles to each flank. Clinical photographs show flank fat reduction between baseline (a, c) and 12 weeks post-treatment (b, d). This subject lost only 0.9 lbs. from her baseline visit to her final photographic evaluation, 12 weeks later.