Experimental verification of SNR and parallel imaging improvements using composite arrays.

Adam Maunder1, B Gino Fallone, Mojgan Daneshmand, Nicola De Zanche.   

Abstract

Composite MRI arrays consist of triplets where two orthogonal upright loops are placed over the same imaging area as a standard surface coil. The optimal height of the upright coils is approximately half the width for the 7 cm coils used in this work. Resistive and magnetic coupling is shown to be negligible within each coil triplet. Experimental evaluation of imaging performance was carried out on a Philips 3 T Achieva scanner using an eight-coil composite array consisting of three surface coils and five upright loops, as well as an array of eight surface coils for comparison. The composite array offers lower overall coupling than the traditional array. The sensitivities of upright coils are complementary to those of the surface coils and therefore provide SNR gains in regions where surface coil sensitivity is low, and additional spatial information for improved parallel imaging performance. Near the surface of the phantom the eight-channel surface coil array provides higher overall SNR than the composite array, but this advantage disappears beyond a depth of approximately one coil diameter, where it is typically more challenging to improve SNR. Furthermore, parallel imaging performance is better with the composite array compared with the surface coil array, especially at high accelerations and in locations deep in the phantom. Composite arrays offer an attractive means of improving imaging performance and channel density without reducing the size, and therefore the loading regime, of surface coil elements. Additional advantages of composite arrays include minimal SNR loss using root-sum-of-squares combination compared with optimal, and the ability to switch from high to low channel density by merely selecting only the surface elements, unlike surface coil arrays, which require additional hardware.
Copyright © 2014 John Wiley & Sons, Ltd.

Keywords:  3D coils; RF coil; RF coil arrays; SNR; array combination; composite arrays; parallel imaging; sample loading

Mesh:

Year:  2014        PMID: 25388793     DOI: 10.1002/nbm.3230

Source DB:  PubMed          Journal:  NMR Biomed        ISSN: 0952-3480            Impact factor:   4.044


  3 in total

1.  Partially orthogonal resonators for magnetic resonance imaging.

Authors:  Jorge Chacon-Caldera; Matthias Malzacher; Lothar R Schad
Journal:  Sci Rep       Date:  2017-02-10       Impact factor: 4.379

2.  Comparison of MEMS switches and PIN diodes for switched dual tuned RF coils.

Authors:  Adam Maunder; Madhwesha Rao; Fraser Robb; Jim M Wild
Journal:  Magn Reson Med       Date:  2018-03-09       Impact factor: 4.668

3.  Low-Noise Active Decoupling Circuit and its Application to 13C Cryogenic RF Coils at 3 T.

Authors:  Juan Diego Sanchez-Heredia; Esben Søvsø Szocska Hansen; Christoffer Laustsen; Vitaliy Zhurbenko; Jan Henrik Ardenkjær-Larsen
Journal:  Tomography       Date:  2017-03
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.