Literature DB >> 25379678

Risk assessment of component failure modes and human errors using a new FMECA approach: application in the safety analysis of HDR brachytherapy.

M Giardina1, F Castiglia, E Tomarchio.   

Abstract

Failure mode, effects and criticality analysis (FMECA) is a safety technique extensively used in many different industrial fields to identify and prevent potential failures. In the application of traditional FMECA, the risk priority number (RPN) is determined to rank the failure modes; however, the method has been criticised for having several weaknesses. Moreover, it is unable to adequately deal with human errors or negligence. In this paper, a new versatile fuzzy rule-based assessment model is proposed to evaluate the RPN index to rank both component failure and human error. The proposed methodology is applied to potential radiological over-exposure of patients during high-dose-rate brachytherapy treatments. The critical analysis of the results can provide recommendations and suggestions regarding safety provisions for the equipment and procedures required to reduce the occurrence of accidental events.

Entities:  

Mesh:

Year:  2014        PMID: 25379678     DOI: 10.1088/0952-4746/34/4/891

Source DB:  PubMed          Journal:  J Radiol Prot        ISSN: 0952-4746            Impact factor:   1.394


  4 in total

1.  The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management.

Authors:  M Saiful Huq; Benedick A Fraass; Peter B Dunscombe; John P Gibbons; Geoffrey S Ibbott; Arno J Mundt; Sasa Mutic; Jatinder R Palta; Frank Rath; Bruce R Thomadsen; Jeffrey F Williamson; Ellen D Yorke
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

2.  Strategies for effective physics plan and chart review in radiation therapy: Report of AAPM Task Group 275.

Authors:  Eric Ford; Leigh Conroy; Lei Dong; Luis Fong de Los Santos; Anne Greener; Grace Gwe-Ya Kim; Jennifer Johnson; Perry Johnson; James G Mechalakos; Brian Napolitano; Stephanie Parker; Deborah Schofield; Koren Smith; Ellen Yorke; Michelle Wells
Journal:  Med Phys       Date:  2020-04-15       Impact factor: 4.071

3.  Electromagnetic tracking (EMT) technology for improved treatment quality assurance in interstitial brachytherapy.

Authors:  Markus Kellermeier; Jens Herbolzheimer; Stephan Kreppner; Michael Lotter; Vratislav Strnad; Christoph Bert
Journal:  J Appl Clin Med Phys       Date:  2017-01       Impact factor: 2.102

4.  Failure mode effect and criticality analysis of ultrasound device by classification tracking.

Authors:  Longchen Wang; Bin Li; Bing Hu; Guofeng Shen; Yunxin Zheng; Yuanyi Zheng
Journal:  BMC Health Serv Res       Date:  2022-04-01       Impact factor: 2.655

  4 in total