Literature DB >> 25379660

Machine-driven versus manual insertion mode: influence on primary stability of orthodontic mini-implants.

Domen Novsak1, Magda Trinajstic Zrinski, Stjepan Spalj.   

Abstract

PURPOSE: The study aimed to explore the effect of the insertion method on maximal insertion torque as a measure of primary stability while controlling for the effect of cortical bone thickness, mini-implant length and diameter, and vertical insertion force on insertion torque.
METHODS: Six types of mini-implants (Dual Top; Jeil Medical, Corp.) with diameters of 1.4, 1.6, and 2.0 mm and lengths of 6 and 8 mm were inserted manually and in a machine-driven mode into pig rib bone samples, and experiments were repeated 10 times, which totaled 120 tested implants in 120 pig rib samples. Cortical bone thickness was measured with a sliding caliper, whereas insertion torque and vertical insertion forces were recorded with a specially designed device.
RESULTS: Significant predictors of better primary stability are thicker cortical bone (explaining 24.2% of variability), wider diameter (20.6%), manual insertion (9.9%), greater length (3.7%), higher maximal vertical insertion force (2.2%), and lower vertical force at maximal insertion torque (1.4%).
CONCLUSIONS: Manual insertion is associated with higher primary stability of orthodontic mini-implants than mechanical insertion, but thicker cortical bone and larger implant diameter seem to be stronger predictors of primary stability.

Entities:  

Mesh:

Year:  2015        PMID: 25379660     DOI: 10.1097/ID.0000000000000174

Source DB:  PubMed          Journal:  Implant Dent        ISSN: 1056-6163            Impact factor:   2.454


  3 in total

1.  Er:YAG laser, piezosurgery, and surgical drill for bone decortication during orthodontic mini-implant insertion: primary stability analysis-an animal study.

Authors:  Jacek Matys; Rafał Flieger; Gianluca Tenore; Kinga Grzech-Leśniak; Umberto Romeo; Marzena Dominiak
Journal:  Lasers Med Sci       Date:  2017-11-10       Impact factor: 3.161

2.  Marginal bone resorption of posterior mandible dental implants with different insertion methods.

Authors:  Ehsan Aliabadi; Saeid Tavanafar; Mohammad Saleh Khaghaninejad
Journal:  BMC Oral Health       Date:  2020-01-31       Impact factor: 2.757

Review 3.  Revisiting the Complications of Orthodontic Miniscrew.

Authors:  Van Mai Truong; Soyeon Kim; Jaeheon Kim; Joo Won Lee; Young-Seok Park
Journal:  Biomed Res Int       Date:  2022-08-01       Impact factor: 3.246