Literature DB >> 25376042

Design of Endoscopic Capsule With Multiple Cameras.

Yingke Gu, Xiang Xie, Guolin Li, Tianjia Sun, Dan Wang, Zheng Yin, Pengfei Zhang, Zhihua Wang.   

Abstract

In order to reduce the miss rate of the wireless capsule endoscopy, in this paper, we propose a new system of the endoscopic capsule with multiple cameras. A master-slave architecture, including an efficient bus architecture and a four level clock management architecture, is applied for the Multiple Cameras Endoscopic Capsule (MCEC). For covering more area of the gastrointestinal tract wall with low power, multiple cameras with a smart image capture strategy, including movement sensitive control and camera selection, are used in the MCEC. To reduce the data transfer bandwidth and power consumption to prolong the MCEC's working life, a low complexity image compressor with PSNR 40.7 dB and compression rate 86% is implemented. A chipset is designed and implemented for the MCEC and a six cameras endoscopic capsule prototype is implemented by using the chipset. With the smart image capture strategy, the coverage rate of the MCEC prototype can achieve 98% and its power consumption is only about 7.1 mW.

Mesh:

Year:  2014        PMID: 25376042     DOI: 10.1109/TBCAS.2014.2359012

Source DB:  PubMed          Journal:  IEEE Trans Biomed Circuits Syst        ISSN: 1932-4545            Impact factor:   3.833


  2 in total

Review 1.  Wireless endoscopy in 2020: Will it still be a capsule?

Authors:  Anastasios Koulaouzidis; Dimitris K Iakovidis; Alexandros Karargyris; Emanuele Rondonotti
Journal:  World J Gastroenterol       Date:  2015-05-07       Impact factor: 5.742

2.  3D DCT Based Image Compression Method for the Medical Endoscopic Application.

Authors:  Jiawen Xue; Li Yin; Zehua Lan; Mingzhu Long; Guolin Li; Zhihua Wang; Xiang Xie
Journal:  Sensors (Basel)       Date:  2021-03-05       Impact factor: 3.576

  2 in total