Literature DB >> 25367866

Thirteen-year outcomes in the Anatomique Benoist Girard II hip prosthesis.

Michael J M Catanach1, Rami M Sorial, Guy D Eslick.   

Abstract

BACKGROUND: The Stryker Anatomique Benoist Girard (ABG) hip implant system was a commonly used cementless prosthesis in the early 2000s, which fell from favour after several studies emerged implicating the prosthesis in high rates of revision. This retrospective, single-surgeon clinical study examines the fracture rate, revision rate and reasons for revision in 500 consecutive ABG II primary total conventional hip replacements.
METHODS: Follow-up was conducted by audit of patient notes, patient mailout survey, patient phone contact and audit of the Australian National Joint Replacement Registry (NJRR) database to find instances of fracture and revision. End points were periprosthetic fracture and revision for any reason.
RESULTS: Follow-up was 1.2-13.8 years with a mean of 6.58 years. Of the 500 hips, 17 (3.4%) had undergone a revision. Of these, 13 were due to periprosthetic fracture. Four further fractures occurred that were not revised. Eight of these periprosthetic fractures occurred within 1 year post-operatively. There were four revisions for recurrent dislocations. Kaplan-Meier survival curve demonstrates a 93.7% survivorship at 6.58 years for revision for any reason. Multivariate analysis showed the only statistically significant factor for increased risk of revision was smaller stem size.
CONCLUSION: Our results were consistent with the literature in that the ABG II system has good medium-term results but is prone to periprosthetic fractures, especially in the early post-operative period. When used as a primary total hip arthroplasty, the ABG II system has an 8-year revision rate of 5.6% compared with the 4.9% of all primary total hip arthroplasties.
© 2014 Royal Australasian College of Surgeons.

Entities:  

Keywords:  ABG; arthroplasty; hip; revision; total

Mesh:

Year:  2014        PMID: 25367866     DOI: 10.1111/ans.12894

Source DB:  PubMed          Journal:  ANZ J Surg        ISSN: 1445-1433            Impact factor:   1.872


  3 in total

1.  Good stability of a cementless, anatomically designed femoral stem in aging women: a 9-year RSA study of 32 patients.

Authors:  Erik Aro; Jessica J Alm; Niko Moritz; Kimmo Mattila; Hannu T Aro
Journal:  Acta Orthop       Date:  2018-07-10       Impact factor: 3.717

Review 2.  Periprosthetic fracture as a late mode of failure of the Anatomique Benoist Girard II femoral prosthesis.

Authors:  Jonathan S Mulford; Ronnie Mathew; David Penn; Alana R Cuthbert; Richard De Steiger
Journal:  ANZ J Surg       Date:  2022-02-21       Impact factor: 2.025

3.  Clinical results at 10 years of minimum follow-up with the ABG 2 hip arthroplasty, matched with ceramic-on-ceramic bearings.

Authors:  Remy Coulomb; Jad Mansour; Jérome Essig; Gérard Asencio; Pascal Kouyoumdjian
Journal:  SICOT J       Date:  2022-08-15
  3 in total