Abstract

Background Scaphoid excision and limited intercarpal fusion is a common surgical procedure performed for degenerative disorders of the wrist including scapholunate advanced collapse (SLAC) and scaphoid nonunion advanced collapse (SNAC) wrist deformities. Postoperative screw migration is a rare but devastating complication that can result in severe degenerative changes in the radiocarpal joint. Questions/Purposes The purpose of this study is to report on a series of patients who developed proximal migration of their hardware following limited intercarpal fusions with headless compression screws. Patients and Methods Four patients were identified between 2001 and 2012 who were indicated for and underwent scaphoid excision and midcarpal fusions with headless compression screw fixation and subsequently developed hardware migration with screw protrusion into the radiocarpal joint. Detailed chart review was performed. Results Mean age at surgery was 64 years (57-69 years). All patients had the diagnosis of SLAC wrist. Mean time to detection of failure was 6 months (4-8 months). All patients demonstrated radiographic union prior to failure based on plain films. Radiographs revealed screw backout with erosion of the radial lunate facet in all patients. Calculated carpal height ratios demonstrated a drop from an average 44.2% to 39.5% at the time of hardware migration. All four patients underwent hardware removal. One patient was not indicated for any further surgery, and two patients underwent further revision surgery. All three patients reported complete pain relief. One patient refused a salvage procedure and had subsequent persistent pain. Conclusions This study reports a serious complication of scaphoid excision and midcarpal fusion performed with headless compression screws. We advise surgeons to be aware of this potential complication and consider employing methods to reduce the risk of hardware migration. Additionally, we recommend at least 8 months of clinical and radiographic follow-up postoperatively to enable early intervention if necessary. Level of Evidence Level IV, therapeutic study.

Entities:  

Keywords:  SLAC wrist; hardware failure; headless compression screws

Year:  2014        PMID: 25364638      PMCID: PMC4208965          DOI: 10.1055/s-0034-1384750

Source DB:  PubMed          Journal:  J Wrist Surg        ISSN: 2163-3916


  15 in total

1.  Complications of circular plate fixation for four-corner arthrodesis.

Authors:  M K Shindle; K J Burton; A J Weiland; B G Domb; S W Wolfe
Journal:  J Hand Surg Eur Vol       Date:  2006-10-17

2.  A prospective outcomes study of four-corner wrist arthrodesis using a circular limited wrist fusion plate for stage II scapholunate advanced collapse wrist deformity.

Authors:  Kevin C Chung; Andrew J Watt; Sandra V Kotsis
Journal:  Plast Reconstr Surg       Date:  2006-08       Impact factor: 4.730

3.  Complications and outcome of four-corner arthrodesis: circular plate fixation versus traditional techniques.

Authors:  Michael C Vance; Jon D Hernandez; Michael L Didonna; Peter J Stern
Journal:  J Hand Surg Am       Date:  2005-11       Impact factor: 2.230

4.  Four-corner fusion and scaphoid excision using headless compression screws for SLAC and SNAC wrist deformities.

Authors:  Allison Alexander Richards; Ahmed M Afifi; Moheb S Moneim
Journal:  Tech Hand Up Extrem Surg       Date:  2011-06

5.  Scaphoid excision and 4-corner fusion using retrograde headless compression screws.

Authors:  Brandon Ball; Joseph W Bergman
Journal:  Tech Hand Up Extrem Surg       Date:  2012-12

6.  Scapholunate advanced collapse wrist salvage.

Authors:  D Ashmead; H K Watson; C Damon; S Herber; W Paly
Journal:  J Hand Surg Am       Date:  1994-09       Impact factor: 2.230

7.  Functional outcome of four-corner arthrodesis for treatment of grade IV scaphoid non-union.

Authors:  Hani El-Mowafi; Mahmoud El-Hadidi; George W Boghdady; Ehab Y Hasanein
Journal:  Acta Orthop Belg       Date:  2007-10       Impact factor: 0.500

8.  Four-corner arthrodesis using a circular plate and distal radius bone grafting: a consecutive case series.

Authors:  Gregory A Merrell; Erin M McDermott; Arnold-Peter C Weiss
Journal:  J Hand Surg Am       Date:  2008 May-Jun       Impact factor: 2.230

9.  Spider plate fixation: no significant improvement in limited wrist arthrodesis.

Authors:  Evan D Collins; Jose Nolla
Journal:  Tech Hand Up Extrem Surg       Date:  2008-06

10.  Clinical outcomes of scaphoid and triquetral excision with capitolunate arthrodesis versus scaphoid excision and four-corner arthrodesis.

Authors:  R Glenn Gaston; Jeffrey A Greenberg; Robert M Baltera; Alex Mih; Hill Hastings
Journal:  J Hand Surg Am       Date:  2009-09-06       Impact factor: 2.230

View more
  1 in total

1.  An Unusual and Complicated Course of a Giant Cell Tumor of the Capitate Bone.

Authors:  Ingo Schmidt
Journal:  Case Rep Orthop       Date:  2016-10-25
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.