Literature DB >> 25359434

Utility of CT-compatible EEG electrodes in critically ill children.

Nicholas S Abend1, Dennis J Dlugos, Xiaowei Zhu, Erin S Schwartz.   

Abstract

BACKGROUND: Electroencephalographic monitoring is being used with increasing frequency in critically ill children who may require frequent and sometimes urgent brain CT scans. Standard metallic disk EEG electrodes commonly produce substantial imaging artifact, and they must be removed and later reapplied when CT scans are indicated.
OBJECTIVE: To determine whether conductive plastic electrodes caused artifact that limited CT interpretation.
MATERIAL AND METHODS: We describe a retrospective cohort of 13 consecutive critically ill children who underwent 17 CT scans with conductive plastic electrodes during 1 year. CT images were evaluated by a pediatric neuroradiologist for artifact presence, type and severity.
RESULTS: All CT scans had excellent quality images without artifact that impaired CT interpretation except for one scan in which improper wire placement resulted in artifact.
CONCLUSION: Conductive plastic electrodes do not cause artifact limiting CT scan interpretation and may be used in critically ill children to permit concurrent electroencephalographic monitoring and CT imaging.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25359434     DOI: 10.1007/s00247-014-3208-5

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  8 in total

1.  Usefulness of a 1.5 T MRI-compatible EEG electrode system for routine use in the intensive care unit of a tertiary care hospital.

Authors:  Seyed M Mirsattari; Corrine Davies-Schinkel; G Bryan Young; Michael D Sharpe; John R Ives; Donald H Lee
Journal:  Epilepsy Res       Date:  2009-03       Impact factor: 3.045

2.  Imaging compatible electrodes for continuous electroencephalogram monitoring in the intensive care unit.

Authors:  Serge Vulliemoz; Stephen Perrig; Daniel Pellise; Maria I Vargas; Yvan Gasche; John R Ives; Margitta Seeck
Journal:  J Clin Neurophysiol       Date:  2009-08       Impact factor: 2.177

3.  The utility of conductive plastic electrodes in prolonged ICU EEG monitoring.

Authors:  Rohit R Das; Brendan P Lucey; Sherry H-Y Chou; Patricio S Espinosa; Amir A Zamani; Barbara A Dworetzky; Edward B Bromfield; Jong Woo Lee
Journal:  Neurocrit Care       Date:  2008-09-20       Impact factor: 3.210

4.  Electrographic status epilepticus and long-term outcome in critically ill children.

Authors:  Katherine L Wagenman; Taylor P Blake; Sarah M Sanchez; Maria T Schultheis; Jerilynn Radcliffe; Robert A Berg; Dennis J Dlugos; Alexis A Topjian; Nicholas S Abend
Journal:  Neurology       Date:  2014-01-02       Impact factor: 9.910

Review 5.  Pediatric ICU EEG monitoring: current resources and practice in the United States and Canada.

Authors:  Sarah M Sanchez; Jessica Carpenter; Kevin E Chapman; Dennis J Dlugos; William B Gallentine; Christopher C Giza; Joshua L Goldstein; Cecil D Hahn; Sudha K Kessler; Tobias Loddenkemper; James J Riviello; Nicholas S Abend
Journal:  J Clin Neurophysiol       Date:  2013-04       Impact factor: 2.177

6.  Electrographic seizures in pediatric ICU patients: cohort study of risk factors and mortality.

Authors:  Nicholas S Abend; Daniel H Arndt; Jessica L Carpenter; Kevin E Chapman; Karen M Cornett; William B Gallentine; Christopher C Giza; Joshua L Goldstein; Cecil D Hahn; Jason T Lerner; Tobias Loddenkemper; Joyce H Matsumoto; Kristin McBain; Kendall B Nash; Eric Payne; Sarah M Sánchez; Iván Sánchez Fernández; Justine Shults; Korwyn Williams; Amy Yang; Dennis J Dlugos
Journal:  Neurology       Date:  2013-06-21       Impact factor: 9.910

7.  Impact of continuous EEG monitoring on clinical management in critically ill children.

Authors:  Nicholas S Abend; Alexis A Topjian; Ana M Gutierrez-Colina; Maureen Donnelly; Robert R Clancy; Dennis J Dlugos
Journal:  Neurocrit Care       Date:  2011-08       Impact factor: 3.210

8.  Nonconvulsive seizures are common in critically ill children.

Authors:  N S Abend; A M Gutierrez-Colina; A A Topjian; H Zhao; R Guo; M Donnelly; R R Clancy; D J Dlugos
Journal:  Neurology       Date:  2011-02-09       Impact factor: 9.910

  8 in total
  1 in total

Review 1.  Electroencephalogram in the intensive care unit: a focused look at acute brain injury.

Authors:  Ayham Alkhachroum; Brian Appavu; Benjamin Rohaut; Jan Claassen; Satoshi Egawa; Brandon Foreman; Nicolas Gaspard; Emily J Gilmore; Lawrence J Hirsch; Pedro Kurtz; Virginie Lambrecq; Julie Kromm; Paul Vespa; Sahar F Zafar
Journal:  Intensive Care Med       Date:  2022-08-23       Impact factor: 41.787

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.