Literature DB >> 25346136

Defining optical distortion in rigid endoscopes.

Eric Abel1, Nikolaos Fotiadis, Mohammed Miah, Paul White.   

Abstract

OBJECTIVES/HYPOTHESIS: To describe lens and perspective distortion using new measures that have practical meaning to the surgeon, and to apply these measures to show the extent of optical distortion in rigid endoscopes used in endoscopic sinus surgery. STUDY
DESIGN: Laboratory measurements on rigid endoscopes.
METHODS: Barrel and perspective distortion were measured in 4-mm diameter 0°, 30°, 45°, and 70° rigid sinus endoscopes. Images of square grids were obtained with the endoscopes aligned in a specially constructed test rig. The terms relative size (RS) and relative distance (RD) were introduced to describe size and distance errors; and the term relative angle (RA) was used for assessing perspective errors.
RESULTS: All the endoscopes exhibited similar barrel distortion. RS of the image at the periphery was 52%; RD was 80%. For RA values of 30°, 45°, and 70°, RS values were 77%, 58%, and 32%, respectively. Objects at the edge of the surgical field appear significantly more distant than suggested by their screen position. Perspective distortion occurs, unless RA = 0°. Barrel distortion of the lens helped to offset the effects of perspective distortion.
CONCLUSION: Optical distortion can be quantified and understood using straightforward definitions. High levels of distortion are common, particularly due to perspective distortion, which is dependent on RA but independent of barrel distortion and the viewing angle of the endoscope.
© 2014 The American Laryngological, Rhinological and Otological Society, Inc.

Entities:  

Keywords:  Rigid endoscopes; barrel distortion; definitions of distortion; depth perception; perspective distortion

Mesh:

Year:  2014        PMID: 25346136     DOI: 10.1002/lary.24971

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  2 in total

1.  Is 3D faster and safer than 4K laparoscopic cholecystectomy? A randomised-controlled trial.

Authors:  Matt Dunstan; Ralph Smith; Katie Schwab; Andrea Scala; Piers Gatenby; Martin Whyte; Tim Rockall; Iain Jourdan
Journal:  Surg Endosc       Date:  2019-07-18       Impact factor: 4.584

2.  Light intensity distribution in images from rigid endoscopes used in minimal access sinus surgery.

Authors:  Eric W Abel; Nikolaos Fotiadis; Paul S White
Journal:  Laryngoscope Investig Otolaryngol       Date:  2021-11-23
  2 in total