Literature DB >> 25344591

Surgical fires in laser laryngeal surgery: are we safe enough?

Soham Roy1, Lee P Smith2.   

Abstract

OBJECTIVE: Laser surgery of the larynx and airway remains high risk for the formation of operating room fire. Traditional methods of fire prevention have included use of "laser safe" tubes, inflation of a protective cuff with saline, and wet pledgets to protect the endotracheal tube from laser strikes. We tested a mechanical model of laser laryngeal surgery to evaluate the fire risk. STUDY
DESIGN: Mechanical model.
SETTING: Laboratory. SUBJECTS AND METHODS: An intubation mannequin was positioned for suspension microlaryngoscopy. A Laser-Shield II cuffed endotracheal tube was placed through the larynx and the cuff inflated using saline. Wet pledgets covered the inflated cuff. A CO2 laser created an inadvertent cuff strike at varying oxygen concentrations. Risk reduction measures were implemented to discern any notable change in the outcome after fire.
RESULTS: At 100% FiO2 an immediate fire with sustained flame was created and at 40% FiO2 a near immediate sustained flame was created. At 29% FiO2, a small nonsustained flame was noted. At room air, no fire was created. There was no discernible difference in the severity of laryngeal damage after the fire occurred whether the tube was immediately pulled from the mannequin or if saline was poured down the airway as a first response.
CONCLUSIONS: While "laser safe" tubes provide a layer of protection against fires, they are not fire proof. Inadvertent cuff perforation may result in fire formation in low-level oxygen enriched environments. Placement of wet pledgets do not provide absolute protection. Endotracheal tube (ETT) cuffs should be placed distally well away from an inadvertent laser strike while maintaining the minimum supplemental oxygen necessary. © American Academy of Otolaryngology—Head and Neck Surgery Foundation 2014.

Entities:  

Keywords:  fire; laryngeal surgery; laser; safety

Mesh:

Substances:

Year:  2014        PMID: 25344591     DOI: 10.1177/0194599814555853

Source DB:  PubMed          Journal:  Otolaryngol Head Neck Surg        ISSN: 0194-5998            Impact factor:   3.497


  6 in total

1.  The Importance of State and Context in Safe Interoperable Medical Systems.

Authors:  Sandy Weininger; Michael B Jaffe; Michael Robkin; Tracy Rausch; David Arney; Julian M Goldman
Journal:  IEEE J Transl Eng Health Med       Date:  2016-08-08       Impact factor: 3.316

Review 2.  The anesthesia of trachea and bronchus surgery.

Authors:  Zehra Hatipoglu; Mediha Turktan; Alper Avci
Journal:  J Thorac Dis       Date:  2016-11       Impact factor: 2.895

3.  Fire safety study on high-flow nasal oxygen in shared-airway surgeries with diathermy and laser: simulation based on a physical model.

Authors:  Man-Yun Chang; Jui-Hung Chen; Shih-Pin Lin; Wei-Nung Teng; Shu-Wei Liao; Chien-Kun Ting; Mei-Yung Tsou; Hui-Hua Kenny Chiang; Fu-Wei Su
Journal:  J Clin Monit Comput       Date:  2021-03-30       Impact factor: 1.977

4.  Risk of airway fire with the use of KTP laser and high flow humidified oxygen delivery in a laryngeal surgery model.

Authors:  Lucy Huang; Adam Badenoch; Marthinus Vermeulen; Shahid Ullah; Charmaine Woods; Theodore Athanasiadis; Eng Hooi Ooi
Journal:  Sci Rep       Date:  2022-01-11       Impact factor: 4.379

5.  The i-gel supraglottic airway device improves airway management during endobronchial ablative therapy under general anesthesia: a case report.

Authors:  Mee Young Chung; SungJin Hong; MinJung Shin; SeungHee Cha; JiYung Lee
Journal:  J Int Med Res       Date:  2022-08       Impact factor: 1.573

6.  Humidified rapid-insufflation ventilatory exchange is a means of oxygenation during rigid bronchoscopy: A case series.

Authors:  Anna Kornafeld; Sebastian Fernandez-Bussy; David Abia-Trujillo; Juan C Garcia; Ryan M Chadha
Journal:  Respirol Case Rep       Date:  2022-01-25
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.