| Literature DB >> 25336403 |
Ramesh Ramji1, Cheong Fook Cheong, Hiroaki Hirata, Abdur Rub Abdur Rahman, Chwee Teck Lim.
Abstract
The number of receptors expressed by cells plays an important role in controlling cell signaling events, thus determining its behaviour, state and fate. Current methods of quantifying receptors on cells are either laborious or do not maintain the cells in their native form. Here, a method integrating highly sensitive bioluminescence, high precision microfluidics and small footprint of lensfree optics is developed to quantify cell surface receptors. This method is safe to use, less laborious, and faster than the conventional radiolabelling and near field scanning methods. It is also more sensitive than fluorescence based assays and is ideal for high throughput screening. In quantifying β(1) adrenergic receptors expressed on the surface of H9c2 cardiomyocytes, this method yields receptor numbers from 3.12 × 10(5) to 9.36 × 10(5) receptors/cell which are comparable with current methods. This can serve as a very good platform for rapid quantification of receptor numbers in ligand/drug binding and receptor characterization studies, which is an important part of pharmaceutical and biological research.Entities:
Keywords: cell islands; cell surface receptors; gaussia luciferase; lens-free optics; microfluidics; receptor numbers
Mesh:
Substances:
Year: 2014 PMID: 25336403 DOI: 10.1002/smll.201401674
Source DB: PubMed Journal: Small ISSN: 1613-6810 Impact factor: 13.281