Abstract
The results from numerous clinical studies provide guidance for optimizing outcomes related to RV or biventricular pacing in the pacemaker and ICD populations. (1) Programming algorithms to minimize RV pacing is imperative in patients with dual-chamber pacemakers who have intrinsic AV conduction or intermittent AV conduction block. (2) Dual-chamber ICDs should be avoided in candidates without an indication for bradycardia pacing. (3) Alternate RV septal pacing sites may be considered at the time of pacemaker implantation. (4) Biventricular pacing may be beneficial in some patients with mild LV dysfunction. (5) LV lead placement at the site of latest LV activation is desirable. (6) Programming CRT systems to achieve biventricular/LV pacing >98.5% is important. (7) Protocols for AV and VV optimization in patients with CRT are not recommended after device implantation but may be considered for CRT nonresponders. (8) Novel algorithms to maximize the benefit of CRT are in evolution further.Entities:
Keywords: cardiac; cardiac pacemaker, artificial; resynchronization therapy
Mesh:
Year: 2014 PMID: 25336367 DOI: 10.1161/CIRCEP.114.001360
Source DB: PubMed Journal: Circ Arrhythm Electrophysiol ISSN: 1941-3084