| Literature DB >> 25323675 |
Richard M Kesler1, Elizabeth T Hsiao-Wecksler, Robert W Motl, Rachel E Klaren, Ipek Ensari, Gavin P Horn.
Abstract
To better assess the energy expenditure and exertion of firefighters during simulated firefighting activities, a commercial firefighter self-contained breathing apparatus (SCBA) facepiece was modified to interface with a portable metabolic monitoring device (Cosmed K4b(2)) while still functioning as a positive pressure SCBA air supply. To validate the device, standard National Fire Protection Association 1981 SCBA function tests were conducted and 14 subjects performed variable-workload assessments using all combinations of two test devices (Cosmed K4b(2) and metabolic cart) and two masks (modified SCBA facepiece and stock manufacturer-supplied breath collection). Metabolic data collected with the Cosmed K4b(2) via the modified facepiece were found to be accurate when compared to a ParvoMedics Truemax 2400 metabolic cart (average per cent difference: 4.6%). This modified facepiece design is suitable for use in metabolic studies requiring the utilisation of an SCBA system. Furthermore, the well-established overestimation of oxygen consumption from the Cosmed K4b(2) system was replicated.Entities:
Keywords: firefighting; gas analysis; oxygen uptake; respiration; self-contained breathing apparatus
Mesh:
Year: 2014 PMID: 25323675 DOI: 10.1080/00140139.2014.964783
Source DB: PubMed Journal: Ergonomics ISSN: 0014-0139 Impact factor: 2.778