| Literature DB >> 25323519 |
Muhammad Imran Khan1, Zafar Iqbal, Abad Khan.
Abstract
A highly sensitive, precise, and accurate reversed-phase high performance liquid-chromatography/electrochemical detection method for simultaneous determination of the endogenous free α-lipoic acid and dihydrolipoic acid in biological matrices is presented. The two analytes are extracted from samples with acetonitrile-10% m-phosphoric acid solution(aqueous) (50:50 v/v). To determine the total lipoic acid, samples are treated with tris(2-carboxyethyl)phosphine solution in phosphate buffer: pH 2.5 with 85% o-phosphoric acid prior to deproteination. The two analytes are separated on a C18 (150 × 4.6 mm, 5 μm) analytical column using acetonitrile-50 mM phosphate buffer: pH 2.5 with 85% o-phosphoric acid (35:65 v/v) as the isocratic mobile phase pumped at a flow rate of 2.0 ml/min at the column oven temperature of 35 °C. The column eluents are monitored at a potential of 0.9 V. These analytes are efficiently resolved in <7 min.Entities:
Mesh:
Substances:
Year: 2015 PMID: 25323519 DOI: 10.1007/978-1-4939-1441-8_25
Source DB: PubMed Journal: Methods Mol Biol ISSN: 1064-3745