In vivo testing of a novel adjustable glaucoma drainage device.

Adan Villamarin1, Nikos Stergiopulos1, Stéphane Bigler1, André Mermoud2, Alexandre Moulin3, Sylvain Roy4.   

Abstract

PURPOSE: We report on the in vivo testing of a novel noninvasively adjustable glaucoma drainage device (AGDD), which features an adjustable outflow resistance, and assess the safety and efficiency of this implant.
METHODS: Under general anesthesia, the AGDD was implanted on seven white New Zealand rabbits for a duration of 4 months under a scleral flap in a way analogous to the Ex-PRESS device and set in an operationally closed position. The IOP was measured on a regular basis on the operated and control eyes using a rebound tonometer. Once a month the AGDD was adjusted noninvasively from its fully closed to its fully open position and the resulting pressure drop was measured. The contralateral eye was not operated and served as control. After euthanization, the eyes were collected for histology evaluation.
RESULTS: The mean preoperative IOP was 11.1 ± 2.4 mm Hg. The IOP was significantly lower for the operated eye (6.8 ± 2 mm Hg) compared to the nonoperated eye (13.1 ± 1.6 mm Hg) during the first 8 days after surgery. When opening the AGDD from its fully closed to fully open position, the IOP dropped significantly from 11.2 ± 2.9 to 4.8 ± 0.9 mm Hg (P < 0.05).
CONCLUSIONS: Implanting the AGDD is a safe and uncomplicated surgical procedure. The fluidic resistance was noninvasively adjustable during the postoperative period with the AGDD between its fully closed and fully open positions. Copyright 2014 The Association for Research in Vision and Ophthalmology, Inc.

Entities:  

Keywords:  aqueous humor; aqueous shunt; drainage device; filtering surgery; glaucoma drainage device

Mesh:

Year:  2014        PMID: 25316722     DOI: 10.1167/iovs.14-14563

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  4 in total

1.  Clinical research of EX-PRESS drainage device and modified trabeculectomy combined with intravitreal conbercept treatment for neovascular glaucoma.

Authors:  Peng Li; Qing Liu; Xiao-Min Liu; Li Wang; Zhao-Jiang Du; Qian-Yan Kang
Journal:  Int J Ophthalmol       Date:  2022-02-18       Impact factor: 1.779

2.  eyeWatch™ System Combined with Non-plated Intraorbital Tube Insertion for the Management of Refractory Glaucoma: A Case Series.

Authors:  Efstathios T Detorakis; Adan Villamarin; Sylvain Roy; Stephane Bigler; Georgios Bontzos; Constantinos Stergiopulos; Nikos Stergiopulos
Journal:  J Curr Glaucoma Pract       Date:  2020 May-Aug

3.  Comparison of two prototypes of a magnetically adjustable glaucoma implant in rabbits.

Authors:  Birthe Schwerk; Lisa Harder; Claudia Windhövel; Marion Hewicker-Trautwein; Anna Wagner; Jan-Peter Bach; Lena Carolin Voigt; Ulf Hinze; Boris Chichkov; Heinz Haferkamp; Holger Lubatschowski; Stephan Nikolic; Ingo Nolte
Journal:  PLoS One       Date:  2019-04-11       Impact factor: 3.240

4.  Wound Healing Response After Bleb-Forming Glaucoma Surgery With a SIBS Microshunt in Rabbits.

Authors:  Ralph J S van Mechelen; Jarno E J Wolters; Marjolein Herfs; Christian J F Bertens; Marion Gijbels; Leonard Pinchuk; Theo G M F Gorgels; Henny J M Beckers
Journal:  Transl Vis Sci Technol       Date:  2022-08-01       Impact factor: 3.048

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.