Literature DB >> 25308702

Construction of dose response calibration curves for dicentrics and micronuclei for X radiation in a Serbian population.

J Pajic1, B Rakic2, D Jovicic3, A Milovanovic4.   

Abstract

Biological dosimetry using chromosome damage biomarkers is a valuable dose assessment method in cases of radiation overexposure with or without physical dosimetry data. In order to estimate dose by biodosimetry, any biological dosimetry service have to have its own dose response calibration curve. This paper reveals the results obtained after irradiation of blood samples from fourteen healthy male and female volunteers in order to establish biodosimetry in Serbia and produce dose response calibration curves for dicentrics and micronuclei. Taking into account pooled data from all the donors, the resultant fitted curve for dicentrics is: Ydic=0.0009 (±0.0003)+0.0421 (±0.0042)×D+0.0602 (±0.0022)×D(2); and for micronuclei: Ymn=0.0104 (±0.0015)+0.0824 (±0.0050)×D+0.0189 (±0.0017)×D(2). Following establishment of the dose response curve, a validation experiment was carried out with four blood samples. Applied and estimated doses were in good agreement. On this basis, the results reported here give us confidence to apply both calibration curves for future biological dosimetry requirements in Serbia.
Copyright © 2014 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Biological dosimetry; Calibration curves; Dicentrics; Micronuclei; X radiation

Mesh:

Year:  2014        PMID: 25308702     DOI: 10.1016/j.mrgentox.2014.07.009

Source DB:  PubMed          Journal:  Mutat Res Genet Toxicol Environ Mutagen        ISSN: 1383-5718            Impact factor:   2.873


  4 in total

1.  Meeting radiation dosimetry capacity requirements of population-scale exposures by geostatistical sampling.

Authors:  Peter K Rogan; Eliseos J Mucaki; Ruipeng Lu; Ben C Shirley; Edward Waller; Joan H M Knoll
Journal:  PLoS One       Date:  2020-04-24       Impact factor: 3.240

2.  Comparing Lymphocyte Radiosensitivity of Prostate Cancer Patients with Healthy Donors Using Micronuclei and Chemical Premature Chromosome Condensation Tests.

Authors:  Golfam F; Hashemi B; Haeri A; Nikoofar A
Journal:  J Biomed Phys Eng       Date:  2020-08-01

3.  Development of Dose-Response Calibration Curve for Dicentric Chromosome Induced by X-Rays.

Authors:  Yanti Lusiyanti; Mukh Syaifudin; Tuti Budiantari; Sofiati Purnami; Dwi Ramadhani
Journal:  Genome Integr       Date:  2019-07-19

4.  Establishing a Reference Dose-Response Calibration Curve for Dicentric Chromosome Aberrations to Assess Accidental Radiation Exposure in Saudi Arabia.

Authors:  Ghazi A Alsbeih; Khaled S Al-Hadyan; Najla M Al-Harbi; Sara S Bin Judia; Belal A Moftah
Journal:  Front Public Health       Date:  2020-12-15
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.