Do dual-thread orthodontic mini-implants improve bone/tissue mechanical retention?

Yang-Sung Lin1, Yau-Zen Chang, Jian-Hong Yu, Chun-Li Lin.   

Abstract

PURPOSE: The aim of this study was to understand whether the pitch relationship between micro and macro thread designs with a parametrical relationship in a dual-thread mini-implant can improve primary stability.
MATERIALS AND METHODS: Three types of mini-implants consisting of single-thread (ST) (0.75 mm pitch in whole length), dual-thread A (DTA) with double-start 0.375 mm pitch, and dual-thread B (DTB) with single-start 0.2 mm pitch in upper 2-mm micro thread region for performing insertion and pull-out testing. Histomorphometric analysis was performed in these specimens in evaluating peri-implant bone defects using a non-contact vision measuring system.
RESULTS: The maximum inserted torque (Tmax) in type DTA was found to be the smallest significantly, but corresponding values found no significant difference between ST and DTB. The largest pull-out strength (Fmax) in the DTA mini-implant was found significantly greater than that for the ST mini-implant regardless of implant insertion orientation. Mini-implant engaged the cortical bone well as observed in ST and DTA types.
CONCLUSION: Dual-thread mini-implant with correct micro thread pitch (parametrical relationship with macro thread pitch) in the cortical bone region can improve primary stability and enhanced mechanical retention.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25290278     DOI: 10.1097/ID.0000000000000144

Source DB:  PubMed          Journal:  Implant Dent        ISSN: 1056-6163            Impact factor:   2.454


  2 in total

1.  Developing Customized Dental Miniscrew Surgical Template from Thermoplastic Polymer Material Using Image Superimposition, CAD System, and 3D Printing.

Authors:  Yu-Tzu Wang; Jian-Hong Yu; Lun-Jou Lo; Pin-Hsin Hsu; CHun-Li Lin
Journal:  Biomed Res Int       Date:  2017-02-09       Impact factor: 3.411

2.  Biomechanical comparison of pedicle screw fixation strength in synthetic bones: Effects of screw shape, core/thread profile and cement augmentation.

Authors:  Mu-Yi Liu; Tsung-Ting Tsai; Po-Liang Lai; Ming-Kai Hsieh; Lih-Huei Chen; Ching-Lung Tai
Journal:  PLoS One       Date:  2020-02-21       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.