| Literature DB >> 25282356 |
Kyoko Nakazaki, Shingo Kitamura, Yuki Motomura, Akiko Hida, Yuichi Kamei, Naoki Miura, Kazuo Mishima1.
Abstract
BACKGROUND: This study aimed to develop an algorithm for determining sleep/wake states by using chronological data on the amount of physical activity (activity intensity) measured with the FS-750 actigraph, a device that can be worn at the waist, allows for its data to be downloaded at home, and is suitable for use in both sleep research and remote sleep medicine.Entities:
Mesh:
Year: 2014 PMID: 25282356 PMCID: PMC4203879 DOI: 10.1186/1880-6805-33-31
Source DB: PubMed Journal: J Physiol Anthropol ISSN: 1880-6791 Impact factor: 2.867
Accuracy of sleep/wake determination using the FS-750 actigraph
| Group A a | Group B a |
|
| ||
|---|---|---|---|---|---|
| Agreement rates (%) | Overall | 84.7 ± 3.0 | 85.4 ± 2.8 | -0.157 | 0.876 |
| Stage W | 66.0 ± 6.3 | 64.9 ± 7.1 | 0.114 | 0.910 | |
| Stage 1 | 65.5 ± 6.5 | 65.2 ± 5.0 | 0.034 | 0.973 | |
| Stage 2 | 91.5 ± 2.0 | 93.4 ± 1.2 | -0.791 | 0.435 | |
| Stage 3 + 4 | 99.0 ± 0.6 | 98.2 ± 0.7 | 0.884 | 0.383 | |
| Stage REM | 84.3 ± 5.3 | 82.4 ± 3.7 | 0.303 | 0.764 | |
| Sensitivity (%) | 88.3 ± 2.8 | 90.0 ± 1.5 | -0.522 | 0.605 | |
| Specificity (%) | 66.0 ± 6.3 | 64.9 ± 7.1 | 0.114 | 0.910 |
aValues are expressed as mean ± SE.
Figure 1Optimization of sleep latency determined by the FS-750 (SL ) SLACT is the interval between the time lights were turned off and the time of the first SACT (sleep-onset time) among the sleep states that appeared continuously for more than n epoch for the first time after the lights were turned off. The horizontal axis shows the n defined above. The vertical axis shows sleep latency (min) defined for each n. *P < .05 and **P < .01, significant intraclass correlation between sleep latency determined by polysomnography (SLPSG) and SLACT. †P < .05, significant difference between SLPSG and SLACT (paired t-test). Values are expressed as mean ± SE.
Figure 2Optimization of wake after sleep onset determined by the FS-750 (WASO ). WASOACT is the total time that wake epochs determined by the FS-750 (WACT) appeared continuously for more than n epochs after sleep onset. The horizontal axis shows the n defined above. The vertical axis shows wake after sleep onset (min) defined for each n. *P < .05 and **P < .01, significant intraclass correlation between wake after sleep onset determined by polysomnography (WASOPSG) and WASOACT. †P < .05 and ‡P < .01, significant difference between WASOPSG and WASOACT (paired t-test). Values are expressed as mean ± SE.
Optimized sleep parameters determined by the FS-750 actigraph
| FS-750 | ||||||||
|---|---|---|---|---|---|---|---|---|
| Polysomnography | ||||||||
| Before optimization | After optimization | |||||||
| Mean ± SE | Mean ± SE |
| ICC | Mean ± SE |
| ICC | ||
| Sleep latency | min | 11.8 ± 2.9 | 6.7 ± 1.1 | 1.766 | 0.124 | 12.7 ± 2.7 | -0.310 | 0.403** |
| Wake after sleep onset | min | 44.1 ± 11.5 | 69.5 ± 8.8 | -2.360* | 0.419** | 41.2 ± 7.9 | 0.271 | 0.427** |
| Total sleep time | min | 424.2 ± 13.0 | 403.8 ± 8.7 | 1.716 | 0.411** | 426.1 ± 8.0 | -0.165 | 0.412** |
| Sleep efficiency | % | 88.4 ± 2.7 | 84.1 ± 1.8 | 1.716 | 0.411** | 88.8 ± 1.7 | -0.166 | 0.412** |
*P < .05, **P < .01.
Overall agreement rates, sensitivity, and specificity for sleep/wake states before and after the application of optimization rules
| Before optimization a | After optimization a |
|
| ||
|---|---|---|---|---|---|
| Agreement rates (%) | Overall | 85.0 ± 2.0 | 88.4 ± 2.1 | -6.087 | <0.001 |
| Stage W | 65.4 ± 4.7 | 56.5 ± 5.1 | 2.814 | 0.008 | |
| Stage 1 | 65.4 ± 4.1 | 77.1 ± 3.9 | -5.615 | <0.001 | |
| Stage 2 | 92.4 ± 1.2 | 96.2 ± 0.9 | -7.159 | <0.001 | |
| Stage 3 + 4 | 98.6 ± 0.5 | 99.4 ± 0.4 | -3.128 | 0.004 | |
| Stage REM | 83.3 ± 3.2 | 90.3 ± 2.8 | -6.651 | <0.001 | |
| Sensitivity (%) | 89.1 ± 1.6 | 93.7 ± 1.4 | -8.816 | <0.001 | |
| Specificity (%) | 65.4 ± 4.7 | 56.5 ± 5.1 | 2.814 | 0.008 |
aValues are expressed as mean ± SE (n =34).