Literature DB >> 25281942

Phantomless patient-specific TomoTherapy QA via delivery performance monitoring and a secondary Monte Carlo dose calculation.

Lydia L Handsfield1, Ryan Jones1, David D Wilson1, Jeffery V Siebers1, Paul W Read1, Quan Chen1.   

Abstract

PURPOSE: To describe the validation and implementation of a novel quality assurance (QA) system for TomoTherapy using a Monte Carlo (MC)-based secondary dose calculation and CT detector-based multileaf collimator (MLC) leaf opening time measurement QA verification. This system is capable of detecting plan transfer and delivery errors and evaluating the dosimetric impact of those errors.
METHODS: The authors' QA process, MCLogQA, utilizes an independent pretreatment MC secondary dose calculation and postdelivery TomoTherapy exit detector-based MLC sinogram comparison and log file examination to confirm accurate dose calculation, accurate dose delivery, and to verify machine performance. MC radiation transport simulations are performed to estimate patient dose utilizing prestored treatment machine-specific phase-space information, the patient's planning CT, and MLC sinogram data. Sinogram data are generated from both the treatment planning system (MC_TPS) and from beam delivery log files (MC_Log). TomoTherapy treatment planning dose (DTPS) is compared with DMC_TPS and DMC_Log via dose-volume metrics and mean region of interest dose statistics. For validation, in-phantom ionization chamber dose measurements (DIC) for ten sample patient plans are compared with the computed values.
RESULTS: Dose comparisons to in-phantom ion chamber measurements validate the capability of the MCLogQA method to detect delivery errors. DMC_Log agreed with DIC within 1%, while DTPS values varied by 2%-5% compared to DIC. The authors demonstrated that TomoTherapy treatments can be vulnerable to MLC deviations and interfraction output variations during treatment delivery. Interfractional Linac output variations for each patient were approximately 2% and average output was 1%-1.5% below the gold standard. While average MLC leaf opening time error from patient to patient varied from -0.6% to 1.6%, the MLC leaf errors varied little between fractions for the same patient plan, excluding one patient.
CONCLUSIONS: MCLogQA is a new TomoTherapy QA process that validates the planned dose before delivery and analyzes the delivered dose using the treatment exit detector and log file data. The MCLogQA procedure is an effective and efficient alternative to traditional phantom-based TomoTherapy plan-specific QA because it allows for comprehensive 3D dose verification, accounts for tissue heterogeneity, uses patient CT density tables, reduces total QA time, and provides for a comprehensive QA methodology for each treatment fraction.

Entities:  

Mesh:

Year:  2014        PMID: 25281942     DOI: 10.1118/1.4894721

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Parallel/Opposed: IMRT QA using treatment log files is superior to conventional measurement-based method.

Authors:  Nathan Childress; Quan Chen; Yi Rong
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

2.  Automated customized retrieval of radiotherapy data for clinical trials, audit and research.

Authors:  Marina Romanchikova; Karl Harrison; Neil G Burnet; Andrew Cf Hoole; Michael Pf Sutcliffe; Michael Andrew Parker; Rajesh Jena; Simon James Thomas
Journal:  Br J Radiol       Date:  2018-01-31       Impact factor: 3.039

3.  Measurement-guided volumetric dose reconstruction for helical tomotherapy.

Authors:  Cassandra Stambaugh; Benjamin Nelms; Theresa Wolf; Richard Mueller; Mark Geurts; Daniel Opp; Eduardo Moros; Geoffrey Zhang; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

4.  A novel, yet simple MLC-based 3D-crossfire technique for spatially fractionated GRID therapy treatment of deep-seated bulky tumors.

Authors:  Damodar Pokhrel; Matthew Halfman; Lana Sanford; Quan Chen; Mahesh Kudrimoti
Journal:  J Appl Clin Med Phys       Date:  2020-02-08       Impact factor: 2.102

Review 5.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19

6.  An Automated knowledge-based planning routine for stereotactic body radiotherapy of peripheral lung tumors via DCA-based volumetric modulated arc therapy.

Authors:  Justin Visak; Gary Y Ge; Ronald C McGarry; Marcus Randall; Damodar Pokhrel
Journal:  J Appl Clin Med Phys       Date:  2020-12-03       Impact factor: 2.102

7.  Independent calculation-based verification of volumetric-modulated arc therapy-stereotactic body radiotherapy plans for lung cancer.

Authors:  Tomohiro Ono; Takamasa Mitsuyoshi; Takashi Shintani; Yusuke Tsuruta; Hiraku Iramina; Hideaki Hirashima; Yuki Miyabe; Mitsuhiro Nakamura; Yukinori Matsuo; Takashi Mizowaki
Journal:  J Appl Clin Med Phys       Date:  2020-05-11       Impact factor: 2.102

8.  Leaf open time sinogram (LOTS): a novel approach for patient specific quality assurance of total marrow irradiation.

Authors:  Rajesh Thiyagarajan; Dayananda Shamurailatpam Sharma; Suryakant Kaushik; Mayur Sawant; K Ganapathy; N Arunai Nambi Raj; Srinivas Chilukuri; Sham C Sundar; Kartikeswar Ch Patro; Arjunan Manikandan; M P Noufal; Rangasamy Sivaraman; Jose Easow; Rakesh Jalali
Journal:  Radiat Oncol       Date:  2020-10-14       Impact factor: 3.481