Helical Tomotherapy of the breast: can thermoplastic immobilization improve the reproducibility of the treatment setup and the accuracy of the delivered dose?

S Agostinelli1, S Garelli2, A Bellini2, F Pupillo2, M Guenzi3, D Bosetti3, G Blandino3, G Taccini2.   

Abstract

PURPOSE: To evaluate the impact of thermoplastic mask immobilization in the setup reproducibility and delivered dose for Helical Tomotherapy (HT) of the breast/chest wall.
METHODS: 16 patients treated with Accuray Hi-Art HT for breast-cancer were considered. Patients were positioned supine with arms extended above the head using Civco Wing Board (WB) system. In 50% of patients an Orfit thermoplastic mask was added in order to improve immobilization. Before each treatment fraction a megavoltage CT (MVCT) scan was taken and registered to the planning CT by experienced medical staff. The impact of thermoplastic mask was investigated analysing MVCT shift-roll data and MVCT dose distribution using Planned Adaptive software.
RESULTS: In the analysed cases, the addition of thermoplastic mask had minor impact on the lateral, longitudinal and roll data distribution. Variance of vertical shifts was significantly reduced in the WB + Orfit group. Van Herk's margins were not affected by addition of thermoplastic immobilization. In both groups, target coverage (V95) and maximum dose (D1) were almost identical to planned values. D1 of organs at risk were not significantly different in the two groups.
CONCLUSIONS: Analysis of shift-roll data shows no improvement in the group of patients immobilized with the addition of thermoplastic mask. Van Herk's margin is quite large (7-10 mm) in both groups evidencing the need to perform daily setup correction. The adapted dose distribution complies well with the planned one, showing that if MVCT is used before each treatment fraction, a 3 mm margin (setup component) for CTVs expansion could be adequate.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Breast Tomotherapy; Interfraction error; Thermoplastic mask

Mesh:

Substances:

Year:  2014        PMID: 25281291     DOI: 10.1016/j.ejmp.2014.09.007

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

1.  The heterogeneous CTV-PTV margins should be given for different parts of tumors during tomotherapy.

Authors:  Ying Tong; Guanzhong Gong; Jinhu Chen; Jie Lu; Tonghai Liu; Pinjing Cheng; Yong Yin
Journal:  Oncotarget       Date:  2017-10-06

2.  Comparing Accuracy of Thermoplastic Mask versus Commercial Bra for the Immobilization of Pendulous Breast During Radiation Therapy Treatment: A Retrospective Cohort Study.

Authors:  Zaheeda Mulla; Omar Iskanderani; Amina Weber; Abdullah AlMohamad; Mohammed Sheikh Al-Amoodi; Shamel Soaida
Journal:  Adv Radiat Oncol       Date:  2020-10-14

3.  Cardiac-sparing radiation therapy using positioning breast shell for patients with left-sided breast cancer who are ineligible for breath-hold techniques.

Authors:  Kurian Joseph; Heather Warkentin; Sunita Ghosh; Lee-Anne Polkosnik; Kent Powell; Matthew Brennan; Brad Warkentin; Johanna Jacobs; Khalifa Alkaabi; Susan Chafe; Keith Tankel; Zsolt Gabos; Hong-Wei Liu; Patricia Tai
Journal:  Adv Radiat Oncol       Date:  2017-08-26

4.  TomoMQA: Automated analysis program for MVCT quality assurance of helical tomotherapy.

Authors:  Min Cheol Han; Chae-Seon Hong; Kyung Hwan Chang; Jihun Kim; Su Chul Han; Dong Wook Kim; Kwangwoo Park; Min-Joo Kim; Han Back Shin; Se Hyung Lee; Jin Sung Kim
Journal:  J Appl Clin Med Phys       Date:  2020-04-08       Impact factor: 2.102