Non-invasive cardiac output monitoring in preterm infants undergoing patent ductus arteriosus ligation: a comparison with echocardiography.

Dany E Weisz1, Amish Jain, Joseph Ting, Patrick J McNamara, Afif El-Khuffash.   

Abstract

BACKGROUND: Non-invasive cardiac output monitoring (NICOM; NICOM™) may be useful in the management of extremely premature preterm infants.
OBJECTIVES: To evaluate a new bioreactance-based method of continuous NICOM in preterm infants following patent ductus arteriosus (PDA) ligation.
METHODS: Infants underwent three paired NICOM and echocardiography assessments of stroke volume (SV) and left ventricular output (LVO) in the postoperative period: at 1, 6-8, and 16-18 h postoperatively. NICOM- and echocardiography-measured SV and LVO during those periods were compared using Bland-Altman analysis and the intraclass correlation coefficient (ICC).
RESULTS: Twenty-five infants with a median (interquartile range) gestational age and birth weight of 25.0 weeks (24.5-25.9) and 700 g (615-775), respectively, were included. The overall systematic bias (limits of agreement) across all time points between the NICOM and echocardiography SV readings was 39% (8-69) with NICOM consistently underestimating echocardiography values. There was moderate consistency between NICOM and echocardiography SV values (ICC 0.78, p < 0.001). Compared with the 1-hour scans, the 6- to 8- and 16- to 18-hour scans had increased biases of 7.9% (95% CI 2.5-13.2) and 9.7% (95% CI 3.6-15.8), respectively.
CONCLUSION: Continuous LVO measurement using NICOM was feasible and demonstrated a consistent systematic bias compared with echocardiography in unstable extremely preterm infants without a PDA ligation. NICOM may be used as a trending tool for continuous monitoring in this population, but wide limits of agreement and increasing bias over time suggest it is not interchangeable with echocardiography.
© 2014 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2014        PMID: 25278031     DOI: 10.1159/000365278

Source DB:  PubMed          Journal:  Neonatology        ISSN: 1661-7800            Impact factor:   4.035


  9 in total

1.  Diagnosis and Management of Patent Ductus Arteriosus.

Authors:  Maria Gillam-Krakauer; Jeff Reese
Journal:  Neoreviews       Date:  2018-07

2.  Noninvasive continuous cardiac output and cerebral perfusion monitoring in term infants with neonatal encephalopathy: assessment of feasibility and reliability.

Authors:  Eva Forman; Colm R Breatnach; Stephanie Ryan; Jana Semberova; Jan Miletin; Adrienne Foran; Afif El-Khuffash
Journal:  Pediatr Res       Date:  2017-08-02       Impact factor: 3.756

3.  Electrical velocimetry for non-invasive monitoring of the closure of the ductus arteriosus in preterm infants.

Authors:  Ana Rodríguez Sánchez de la Blanca; M Sánchez Luna; N González Pacheco; M Arriaga Redondo; N Navarro Patiño
Journal:  Eur J Pediatr       Date:  2017-12-08       Impact factor: 3.183

Review 4.  Challenges in Treating Low Blood Pressure in Preterm Infants.

Authors:  Eugene M Dempsey
Journal:  Children (Basel)       Date:  2015-06-15

Review 5.  Neonatal Hemodynamics: From Developmental Physiology to Comprehensive Monitoring.

Authors:  Sabine L Vrancken; Arno F van Heijst; Willem P de Boode
Journal:  Front Pediatr       Date:  2018-04-05       Impact factor: 3.418

Review 6.  Cardiac Output Monitoring in Preterm Infants.

Authors:  Matthew McGovern; Jan Miletin
Journal:  Front Pediatr       Date:  2018-04-03       Impact factor: 3.418

Review 7.  Non-invasive Cardiac Output Monitoring in Neonates.

Authors:  Roisin O'Neill; Eugene M Dempsey; Aisling A Garvey; Christoph E Schwarz
Journal:  Front Pediatr       Date:  2021-01-28       Impact factor: 3.418

8.  Accuracy and Trending Ability of Electrical Biosensing Technology for Non-invasive Cardiac Output Monitoring in Neonates: A Systematic Qualitative Review.

Authors:  Lizelle Van Wyk; Samir Gupta; John Lawrenson; Willem-Pieter de Boode
Journal:  Front Pediatr       Date:  2022-03-17       Impact factor: 3.418

Review 9.  Enhanced Monitoring of the Preterm Infant during Stabilization in the Delivery Room.

Authors:  Daragh Finn; Geraldine B Boylan; C Anthony Ryan; Eugene M Dempsey
Journal:  Front Pediatr       Date:  2016-03-31       Impact factor: 3.418

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.