Literature DB >> 25270636

Comparison of a radiofrequency powered flexible needle with a classic rigid Brockenbrough needle for transseptal punctures in terms of safety and efficacy.

Gaël Jauvert1, Caroline Grimard2, Arnaud Lazarus2, Christine Alonso2.   

Abstract

INTRODUCTION: This study aimed to evaluate the safety and efficacy of utilising an innovative radiofrequency (RF) powered flexible needle to achieve transseptal puncture (TSP). METHODS AND
RESULTS: A RF powered flexible needle (Toronto catheter, Baylis Medical Company Inc.) associated with a stiffer dilator (Torflex Superstrong, Baylis Medical Company Inc.) was used in 125 consecutive patients referred for left sided ablations (mean age=55.6, male=86.5%) and compared with a standard transseptal set (BRK needle, SL0 sheath and dilator, St Jude Medical, Inc.) used in the previous 100 patients (mean age=56, male 82%). TSP was achieved in 95/100 patients in the Brockenbrough group and in all 125 patients in the Toronto group (p=0.01) despite an equivalent proportion of difficult situations (8 and 9% respectively) and patients with a prior TSP (17% vs 24%). 7/100 needle related events (failure, aborted attempt or pericardial effusion) occurred in the Brockenbrough group and none in the Toronto group (p=0.01). The Toronto needle crossed the septum at the first attempt in 123/125 (98.4%) patients and the Brockenbrough needle in 84/95 (88%) patients (p<0.001).
CONCLUSION: Our data suggest that the Toronto RF powered flexible needle is safer and more efficient than a standard Brockenbrough needle and can be used not only in difficult situations but routinely to achieve TSP.
Copyright © 2014 Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Ablations; Safety; Transseptal punctures

Mesh:

Year:  2014        PMID: 25270636     DOI: 10.1016/j.hlc.2014.07.073

Source DB:  PubMed          Journal:  Heart Lung Circ        ISSN: 1443-9506            Impact factor:   2.975