Abstract
The purpose of this study was to evaluate the National Committee for Clinical Laboratory Standards agar dilution, tube dilution, and broth microdilution susceptibility tests for the measurement of teicoplanin MICs. The three standardized tests gave equivalent (within a twofold dilution) results with 98.8 to 99.0% of the 508 gram-positive clinical isolates tested, indicating that either method may be used for teicoplanin MIC determination.Entities:
Mesh:
Substances:
Year: 1989 PMID: 2526821 PMCID: PMC267573 DOI: 10.1128/jcm.27.6.1409-1410.1989
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948