Correction of measured Gamma-Knife output factors for angular dependence of diode detectors and PinPoint ionization chamber.

Hrvoje Hršak1, Marija Majer2, Timor Grego3, Juraj Bibić3, Zdravko Heinrich3.   

Abstract

Dosimetry for Gamma-Knife requires detectors with high spatial resolution and minimal angular dependence of response. Angular dependence and end effect time for p-type silicon detectors (PTW Diode P and Diode E) and PTW PinPoint ionization chamber were measured with Gamma-Knife beams. Weighted angular dependence correction factors were calculated for each detector. The Gamma-Knife output factors were corrected for angular dependence and end effect time. For Gamma-Knife beams angle range of 84°-54°. Diode P shows considerable angular dependence of 9% and 8% for the 18 mm and 14, 8, 4 mm collimator, respectively. For Diode E this dependence is about 4% for all collimators. PinPoint ionization chamber shows angular dependence of less than 3% for 18, 14 and 8 mm helmet and 10% for 4 mm collimator due to volumetric averaging effect in a small photon beam. Corrected output factors for 14 mm helmet are in very good agreement (within ±0.3%) with published data and values recommended by vendor (Elekta AB, Stockholm, Sweden). For the 8 mm collimator diodes are still in good agreement with recommended values (within ±0.6%), while PinPoint gives 3% less value. For the 4 mm helmet Diodes P and E show over-response of 2.8% and 1.8%, respectively. For PinPoint chamber output factor of 4 mm collimator is 25% lower than Elekta value which is generally not consequence of angular dependence, but of volumetric averaging effect and lack of lateral electronic equilibrium. Diodes P and E represent good choice for Gamma-Knife dosimetry.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Correction for angular dependence; Gamma-Knife output factor; PinPoint ionization chamber; Silicon diode

Mesh:

Substances:

Year:  2014        PMID: 25266988     DOI: 10.1016/j.ejmp.2014.09.002

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

1.  Small-field dosimetry with detector-specific output correction factor for single-isocenter stereotactic radiotherapy of single and multiple brain metastases.

Authors:  Tomohiro Ono; Kohei Kawata; Mitsuhiro Nakamura; Megumi Uto; Takashi Mizowaki
Journal:  Radiol Phys Technol       Date:  2022-10-22

2.  A robust measurement point for dose verification in delivery quality assurance for a robotic radiosurgery system.

Authors:  Keita Kurosu; Iori Sumida; Hiroya Shiomi; Hirokazu Mizuno; Hiroko Yamaguchi; Hirofumi Okubo; Keisuke Tamari; Yuji Seo; Osamu Suzuki; Seiichi Ota; Shinichi Inoue; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2017-05-01       Impact factor: 2.724

3.  Assessment of Delivery Quality Assurance for Stereotactic Radiosurgery With Cyberknife.

Authors:  Jun Li; Xile Zhang; Yuxi Pan; Hongqing Zhuang; Junjie Wang; Ruijie Yang
Journal:  Front Oncol       Date:  2021-11-17       Impact factor: 6.244

4.  A novel and effective method for validation and measurement of output factors for Leksell Gamma Knife® Icon™ using TRS 483 protocol.

Authors:  Swapna Lilly Cyriac; Jian Liu; Emel Calugaru; Jenghwa Chang
Journal:  J Appl Clin Med Phys       Date:  2020-09-06       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.