| Literature DB >> 25253788 |
C Pontoizeau1, L Dangers2, V Jarlier3, C E Luyt2, E Guiller4, M H Fievet4, M Lecsö-Bornet5, A Aubry3, F Brossier6.
Abstract
We report here false-positive urinary Legionella pneumophila serogroup 1 and Streptococcus pneumoniae antigen test results due to rabbit antilymphocyte serum treatment and provide a simple and fast solution to rule them out by heating urine.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25253788 PMCID: PMC4313311 DOI: 10.1128/JCM.01916-14