Anderson Freitas1, Welder Fernandes Lula1, Jonathan Sampaio de Oliveira1, Rafael Almeida Maciel1, Diogo Ranier de Macedo Souto1, Patrick Fernandes Godinho1.
Abstract
OBJECTIVE: To analyze the results of biomechanical assays of fixation of Pauwels type III femoral neck fracture in synthetic bone, using 7.5mm cannulated screws in inverted triangle formation, in relation to the control group.Entities:
Keywords: Biomechanical phenomena; Femoral neck fractures; Fracture fixation, internal
Year: 2014 PMID: 25246851 PMCID: PMC4167045 DOI: 10.1590/1413-78522014220400917
Source DB: PubMed Journal: Acta Ortop Bras ISSN: 1413-7852 Impact factor: 0.513
Figure 1A) Synthetic bone Model fixated with three 7.5 mm cannulated screws; B) Radiograph of synthetic models fixated for evaluation of proper positioning of the syntheses.
Figure 2A) Bone model already fixated in the testing machine at pre-test; B) Bone image during the test model at 5mm displacement (Step 1); C) Bone model during the test at 10mm displacement (Step 2).
Figure 3A) Marking for evaluation of pre-test rotational deviation; B) Image of markup misaligned after 1st Step of essay of the test group.
Figure 4(A) Image of synthetic model of the control group on the test machine at pre-test; (B) Image of synthetic model of the control group at post-test.
Figure 5Testing Machine used.
Values from tests on models of the test group in steps 1 and 2.
| Sample | Load with 5 mm displacement (N) | Load with 10 mm displacement (N) | Rotation (degrees) |
|---|---|---|---|
| 1 | 706 | 860 | 3.27 |
| 2 | 514 | 607 | 1.64 |
| 3 | 534 | 614 | 5.22 |
| 4 | 547 | 658 | 6.52 |
| 5 | 594 | 743 | 0,0 |
| Mean | 579 | 696 | 3.33 |
| St. Deviation | 77 | 106 | 2.63 |
Figure 6Force versus displacement curve in models testing of the test group.
Values of assays in samples from the control group.
| Sample | Control group Maximum load (N) |
|---|---|
| 1 | 1544 |
| 2 | 1110 |
| 3 | 1359 |
| 4 | 1194 |
| 5 | 1437 |
| Mean | 1329 |
| St. Deviation | 177 |
Figura 7Curvas força versus deslocamento para o grupo controle.
Figure 8Maximum force at 10 mm displacement according to test group (PC1 board) and control group.