Reproducibility of intravascular ultrasound radiofrequency data analysis (virtual histology) with a 45-MHz rotational imaging catheter in ex vivo human coronary arteries.

Takashi Muramatsu1, Hector M García-García2, Salvatore Brugaletta3, Jung Ho Heo4, Yoshinobu Onuma5, Russell J Fedewa6, Anuja Nair7, Yukio Ozaki8, Patrick W Serruys5.   

Abstract

BACKGROUND: Despite the frequent use of spectral analysis of intravascular ultrasound radiofrequency data (VH(®) IVUS) in clinical studies, the assessment for reproducibility using this with high frequency IVUS remains unexplored.
PURPOSE: The aim of this study was to examine the reproducibility of VH IVUS using 45-MHz rotational IVUS in ex vivo human coronary arteries.
METHODS: Data were collected using 45-MHz VH IVUS (Revolution(®), Volcano Corporation, San Diego, CA, USA) via a series of pullbacks from eight human coronary artery specimens. Imaging data were analyzed by two independent observers. Intraobserver and interobserver reproducibility were assessed using five pullbacks from five vessels. The intercatheter reproducibility was assessed using three different catheters in each of the five vessels. The intracatheter reproducibility was assessed between the two sequential pullbacks from each of the 15 catheters used in the intercatheter assessment.
RESULTS: Geometrical measurements consistently showed low variability (relative difference <10%) and excellent intraclass correlation coefficients (ICCs), ranging from 0.88 to 1.00. With respect to the compositional measurements, the relative differences were predominantly higher than those of geometrical measurements. In particular, fibrous-fatty area showed a higher relative difference (17.5% in intercatheter assessment) compared to fibrous, necrotic core, and dense calcium areas (6.5%, 8.4%, and 6.4%, respectively). However, each compositional measurement also showed acceptable reproducibility (ICCs ranging from 0.82 to 1.00).
CONCLUSIONS: The 45-MHz rotational VH IVUS technology had acceptable reproducibility with respect to geometrical and compositional assessments in ex vivo human coronary arteries. These data are crucial when designing future longitudinal studies addressing geometrical measurements and plaque characterization by 45-MHz VH IVUS.
Copyright © 2014 Japanese College of Cardiology. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Coronary artery disease; Ex vivo; Intravascular ultrasound; Reproducibility; Virtual histology

Mesh:

Year:  2014        PMID: 25242300     DOI: 10.1016/j.jjcc.2014.05.004

Source DB:  PubMed          Journal:  J Cardiol        ISSN: 0914-5087            Impact factor:   3.159


  2 in total

1.  Comparative assessment of "plaque/media" change on three modalities of IVUS immediately after implantation of either everolimus-eluting bioresorbable vascular scaffold or everolimus-eluting metallic stent in Absorb II study.

Authors:  Yaping Zeng; Rafael Cavalcante; Erhan Tenekecioglu; Pannipa Suwannasom; Yohei Sotomi; Carlos Collet; Mahammad Abdelghani; Hans Jonker; Franck Digne; Dieter Horstkotte; Manfred Zehender; Ciro Indolfi; Francesco Saia; Rosario Fiorilli; Bernard Chevalier; Leonardo Bolognese; Javier Goicolea; Shaoping Nie; Yoshinobu Onuma; Patrick W Serruys
Journal:  Int J Cardiovasc Imaging       Date:  2016-12-23       Impact factor: 2.357

2.  Apolipoprotein C3 and necrotic core volume are correlated but also associated with future cardiovascular events.

Authors:  Takayuki Ohwada; Takayuki Sakamoto; Satoshi Suzuki; Yukiko Sugawara; Kazuya Sakamoto; Ayano Ikeda; Fumika Haga; Tomoki Sato; Kazuhiko Nakazato; Yasuchika Takeishi; Kenichi Watanabe
Journal:  Sci Rep       Date:  2022-08-25       Impact factor: 4.996

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.