In vitro assessment of the contact mechanics of reverse-engineered distal humeral hemiarthroplasty prostheses.

Ryan Willing1, Michael Lapner2, Graham J W King2, James A Johnson2.   

Abstract

BACKGROUND: Distal humeral hemiarthroplasty alters cartilage contact mechanics, which may predispose to osteoarthritis. Current prostheses do not replicate the native anatomy, and therefore contribute to these changes. We hypothesized that prostheses reverse-engineered from the native bone shape would provide similar contact patterns as the native articulation.
METHODS: Reverse-engineered hemiarthroplasty prostheses were manufactured for five cadaveric elbows based on CT images of the distal humerus. Passive flexion trials with constant muscle forces were performed with the native articulation intact while bone motions were recorded using a motion tracking system. Motion trials were then repeated after the distal humerus was replaced with a corresponding reverse-engineered prosthesis. Contact areas and patterns were reconstructed using computer models created from CT scan images combined with the motion tracker data. The total contact areas, as well as the contact area within smaller sub-regions of the ulna and radius, were analyzed for changes resulting from hemiarthroplasty using repeated-measures ANOVAs.
FINDINGS: Contact area at the ulna and radius decreased on average 42% (SD 19%, P=.008) and 41% (SD 42%, P=.096), respectively. Contact area decreases were not uniform throughout the different sub-regions, suggesting that contact patterns were also altered.
INTERPRETATION: Reverse-engineered prostheses did not reproduce the same contact pattern as the native joints, possibly because the thickness of the distal humerus cartilage layer was neglected when generating the prosthesis shapes or as a consequence of the increased stiffness of the metallic implants. Alternative design strategies and materials for hemiarthroplasty should be considered in future work.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Contact pattern; Distal humerus; Elbow; Hemiarthroplasty; Reverse-engineered

Mesh:

Year:  2014        PMID: 25238687     DOI: 10.1016/j.clinbiomech.2014.08.015

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  4 in total

1.  Anatomical variations of the trochlear notch angle: MRI analysis of 78 elbows.

Authors:  G Giannicola; M Scacchi; P Sedati; S Gumina
Journal:  Musculoskelet Surg       Date:  2016-11-30

2.  Distal Humeral Trochlear Geometry Associated With the Spatial Variation of the Dynamic Elbow Flexion Axis.

Authors:  Diyang Zou; Xiangjun Hu; Kai-Nan An; Kerong Dai; Xiaowei Yu; Weihua Gong; Tsung-Yuan Tsai
Journal:  Front Bioeng Biotechnol       Date:  2022-06-24

3.  Distal Humerus Morphological Analysis of Chinese Individuals: A Statistical Shape Modeling Approach.

Authors:  Wei Zhao; Yao Guo; Chuangye Xu; Guoxian Pei; Shiva Basnet; Yanjun Pei; Xiuyun Su
Journal:  Orthop Surg       Date:  2022-09-14       Impact factor: 2.279

4.  Comparative Biomechanical Analysis of Stress-Strain State of the Elbow Joint After Displaced Radial Head Fractures.

Authors:  Sergey Strafun; Ievgen Levadnyi; Vasily Makarov; Jan Awrejcewicz
Journal:  J Med Biol Eng       Date:  2017-09-26       Impact factor: 1.553

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.