Early feasibility testing and engineering development of a sutureless beating heart connector for left ventricular assist devices.

Steven C Koenig1, Jorge H Jimenez, Seth D West, Michael A Sobieski, Young Choi, Gretel Monreal, Guruprasad A Giridharan, Kevin G Soucy, Mark S Slaughter.   

Abstract

APK Advanced Medical Technologies (Atlanta, GA) is developing a sutureless beating heart (SBH) left ventricular assist device (LVAD) connector system consisting of anchoring titanium coil, titanium cannula with integrated silicone hemostatic valve, coring and delivery tool, and LVAD locking mechanism to facilitate LVAD inflow surgical procedures. Feasibility testing was completed in human cadavers (n = 4) under simulated normal and hypertensive conditions using saline to observe seal quality in degraded human tissue and assess anatomic fit; acutely in ischemic heart failure bovine model (n = 2) to investigate short-term performance and ease of use; and chronically for 30 days in healthy calves (n = 2) implanted with HeartWare HVAD to evaluate performance and biocompatibility. Complete hemostasis was achieved in human cadavers and animals at LV pressures up to 170 mm Hg. In animals, off-pump (no cardiopulmonary bypass) anchoring of the connector was accomplished in less than 1 minute with no residual bleeding after full delivery and locking of the LVAD; and implant of connector and LVAD were successfully completed in under 10 minutes with total procedure blood loss less than 100 ml. In chronic animals before necropsy, no signs of leakage or disruption at the attachment site were observed at systolic LV pressures >200 mm Hg.

Entities:  

Mesh:

Year:  2014        PMID: 25238500      PMCID: PMC4224637          DOI: 10.1097/MAT.0000000000000134

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  24 in total

1.  Implantation technique for the HeartMate left ventricular assist device.

Authors:  B Radovancevic; O H Frazier; J M Duncan
Journal:  J Card Surg       Date:  1992-09       Impact factor: 1.620

2.  Short-term changes in pulmonary function and respiratory movements after cardiac surgery via median sternotomy.

Authors:  María Ragnarsdóttir; Asdís KristjAnsdóttir; Ingveldur Ingvarsdóttir; Pétur Hannesson; Bjarni Torfason; Lawrence Cahalin
Journal:  Scand Cardiovasc J       Date:  2004-03       Impact factor: 1.589

3.  Hospital costs for left ventricular assist devices for destination therapy: lower costs for implantation in the post-REMATCH era.

Authors:  Leslie W Miller; Karl E Nelson; Robin R Bostic; Kuo Tong; Mark S Slaughter; James W Long
Journal:  J Heart Lung Transplant       Date:  2006-07       Impact factor: 10.247

4.  Off-pump implant of the Jarvik 2000 ventricular assist device through median sternotomy.

Authors:  Ani C Anyanwu; Gregory W Fischer; Irina Plotkina; Sean Pinney; David H Adams
Journal:  Ann Thorac Surg       Date:  2007-10       Impact factor: 4.330

5.  Placement of long-term implantable ventricular assist devices without the use of cardiopulmonary bypass.

Authors:  Benjamin C Sun; Michael S Firstenberg; Louis B Louis; Antonio Panza; Juan A Crestanello; John Sirak; Chittoor B Sai-Sudhakar
Journal:  J Heart Lung Transplant       Date:  2008-06-06       Impact factor: 10.247

6.  Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device.

Authors:  Francis D Pagani; Leslie W Miller; Stuart D Russell; Keith D Aaronson; Ranjit John; Andrew J Boyle; John V Conte; Roberta C Bogaev; Thomas E MacGillivray; Yoshifumi Naka; Donna Mancini; H Todd Massey; Leway Chen; Charles T Klodell; Juan M Aranda; Nader Moazami; Gregory A Ewald; David J Farrar; O Howard Frazier
Journal:  J Am Coll Cardiol       Date:  2009-07-21       Impact factor: 24.094

7.  Advanced heart failure treated with continuous-flow left ventricular assist device.

Authors:  Mark S Slaughter; Joseph G Rogers; Carmelo A Milano; Stuart D Russell; John V Conte; David Feldman; Benjamin Sun; Antone J Tatooles; Reynolds M Delgado; James W Long; Thomas C Wozniak; Waqas Ghumman; David J Farrar; O Howard Frazier
Journal:  N Engl J Med       Date:  2009-11-17       Impact factor: 91.245

8.  Respiratory movements are altered three months and one year following cardiac surgery.

Authors:  Asdís Kristjánsdóttir; María Ragnarsdóttir; Pétur Hannesson; Hans Jakob Beck; Bjarni Torfason
Journal:  Scand Cardiovasc J       Date:  2004-05       Impact factor: 1.589

9.  A less invasive approach to axial flow pump insertion.

Authors:  Igor D Gregoric; Saverio La Francesca; Tim Myers; William Cohn; Pranav Loyalka; Biswajit Kar; Courtney Gemmato; O H Frazier
Journal:  J Heart Lung Transplant       Date:  2008-04       Impact factor: 10.247

10.  First clinical use of the redesigned HeartMate II left ventricular assist system in the United States: a case report.

Authors:  O H Frazier; Reynolds M Delgado; Biswajit Kar; Vijay Patel; Igor D Gregoric; Timothy J Myers
Journal:  Tex Heart Inst J       Date:  2004
View more
  1 in total

1.  Construction and application of service quality evaluation system in the preclinical research on cardiovascular implant devices.

Authors:  Yongchun Cui; Fuliang Luo; Boqing Yang; Bin Li; Qi Zhang; Gopika Das; Guangxin Yue; Jiajie Li; Yue Tang; Xin Wang
Journal:  BMC Med Inform Decis Mak       Date:  2019-02-28       Impact factor: 2.796

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.