Value of peritoneoscopy for nonfunctioning continuous ambulatory peritoneal dialysis catheters.

D W Smith1, R A Rankin.   

Abstract

We examined our results in restoring continuous ambulatory peritoneal dialysis (CAPD) catheter function via peritoneoscopy in 10 procedures performed on nine patients. Malfunction was caused by the catheter being encased in adhesions, being buried in small bowel, or being wrapped and occluded in omentum. In five of these cases we were successful in restoring catheter patency, thereby allowing immediate reinstitution of dialysis. Neither previous abdominal surgeries nor the length of time the catheter had been in place prior to malfunction appeared to be predictive for the cause of malfunction or the success of peritoneoscopic manipulation. We found outpatient peritoneoscopy to be safe and cost effective. We recommend it as the initial procedure to restore patency to malfunctioning CAPD catheters after conservative therapy has failed.

Entities:  

Mesh:

Year:  1989        PMID: 2523835     DOI: 10.1016/s0016-5107(89)72715-2

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  6 in total

Review 1.  Clinical practice guidelines for pediatric peritoneal dialysis.

Authors:  Colin T White; Manjula Gowrishankar; Janusz Feber; Verna Yiu
Journal:  Pediatr Nephrol       Date:  2006-07-04       Impact factor: 3.714

2.  A new laparoscopic technique for the placement of a permanent peritoneal dialysis catheter: the preperitoneal tunneling method.

Authors:  M Comert; A Borazan; E Kulah; B Hamdi Uçan
Journal:  Surg Endosc       Date:  2004-11-11       Impact factor: 4.584

3.  Video-assisted laparoscopic procedures in peritoneal dialysis.

Authors:  S C Jwo; K S Chen; Y Y Lin
Journal:  Surg Endosc       Date:  2003-08-15       Impact factor: 4.584

4.  Laparoscopic correction of peritoneal catheter dysfunction.

Authors:  Gholamhossein Kazemzadeh; Mohammad-Hadi Saeed Modaghegh; Alireza Tavassoli
Journal:  Indian J Surg       Date:  2008-11-26       Impact factor: 0.656

5.  A novel method for salvage of malfunctioning peritoneal dialysis catheter.

Authors:  Ali Akbar Beig; Sayed Mahdi Marashi; Hojatollah Raji Asadabadi; Ali Sharifi; Zohre Nasiri Zarch
Journal:  Urol Ann       Date:  2014-04

6.  Minimally invasive surgery for salvage of malfunctioning peritoneal dialysis catheters.

Authors:  Hrishikesh P Salgaonkar; Ramya Ranjan Behera; Pradeep Chandra Sharma; Avinash Katara; Deepraj S Bhandarkar
Journal:  J Minim Access Surg       Date:  2019 Jan-Mar       Impact factor: 1.407

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.