Ryo Kosaka1, Kazuya Yasui, Masahiro Nishida, Yasuo Kawaguchi, Osamu Maruyama, Takashi Yamane.
Abstract
We have developed a hydrodynamically levitated centrifugal pump as a bridge-to-decision device. The purpose of the present study is to determine the optimal bearing gap of a multiarc radial bearing in the developed blood pump for the reduction of hemolysis. We prepared eight pump models having bearing gaps of 20, 30, 40, 80, 90, 100, 180, and 250 μm. The driving conditions were set to a pressure head of 200 mm Hg and a flow rate of 4 L/min. First, the orbital radius of the impeller was measured for the evaluation of the impeller stability. Second, the hemolytic property was evaluated in an in vitro hemolysis test. As a result, the orbital radius was not greater than 15 μm when the bearing gap was between 20 and 100 μm. The relative normalized index of hemolysis (NIH) ratios in comparison with BPX-80 were 37.67 (gap: 20 μm), 0.95 (gap: 30 μm), 0.96 (gap: 40 μm), 0.82 (gap: 80 μm), 0.77 (gap: 90 μm), 0.92 (gap: 100 μm), 2.76 (gap: 180 μm), and 2.78 (gap: 250 μm). The hemolysis tended to increase at bearing gaps of greater than 100 μm due to impeller instability. When the bearing gap decreased from 30 to 20 μm, the relative NIH ratios increased significantly from 0.95 to 37.67 times (P < 0.01) due to high shear stress. We confirmed that the optimal bearing gap was determined between 30 and 100 μm in the developed blood pump for the reduction of hemolysis.Entities:
Keywords: Centrifugal blood pump; Hemolysis; Hydrodynamic bearing; Impeller trajectory; Radial gap
Mesh:
Year: 2014 PMID: 25234763 DOI: 10.1111/aor.12383
Source DB: PubMed Journal: Artif Organs ISSN: 0160-564X Impact factor: 3.094