Literature DB >> 25234093

System overview of the fully implantable destination therapy--ReinHeart-total artificial heart.

Benedikt Pelletier1, Sotirios Spiliopoulos2, Thomas Finocchiaro3, Felix Graef3, Kristin Kuipers3, Marco Laumen3, Dilek Guersoy2, Ulrich Steinseifer3, Reiner Koerfer2, Gero Tenderich2.   

Abstract

OBJECTIVES: Owing to the lack of suitable allografts, the demand for long-term mechanical circulatory support in patients with biventricular end-stage heart failure is rising. Currently available Total Artificial Heart (TAH) systems consist of pump units with only limited durability, percutaneous tubes and bulky external equipment that limit the quality of life. Therefore we are focusing on the development of a fully implantable, highly durable destination therapy total artificial heart.
METHODS: The ReinHeart-TAH system consists of a passively filling pump unit driven by a low-wear linear drive between two artificial ventricles, an implantable control unit and a compliance chamber. The TAH is powered by a transcutaneous energy transmission system. The flow distribution inside the ventricles was analysed by fluid structure interaction simulation and particle image velocimetry measurements. Along with durability tests, the hydrodynamic performance and flow balance capability were evaluated in a mock circulation loop. Animal trials are ongoing.
RESULTS: Based on fluid structure interaction simulation and particle image velocimetry, blood stagnation areas have been significantly reduced. In the mock circulation loop the ReinHeart-TAH generated a cardiac output of 5 l/min at an operating frequency of 120 bpm and an aortic pressure of 120/80 mmHg. The highly effective preload sensitivity of the passively filling ventricles allowed the sensorless integration of the Frank Starling mechanism. The ReinHeart-TAH effectively replaced the native heart's function in animals for up to 2 days.
CONCLUSIONS: In vitro and in vivo testing showed a safe and effective function of the ReinHeart-TAH system. This has the potential to become an alternative to transplantation. However, before a first-in-man implant, chronic animal trials still have to be completed.
© The Author 2014. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

Entities:  

Keywords:  Fully implantable; Heart failure; Quality of life; Total Artificial Heart TAH

Mesh:

Year:  2014        PMID: 25234093     DOI: 10.1093/ejcts/ezu321

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  6 in total

1.  Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart.

Authors:  Muath Bishawi; Jun-Neng Roan; Jordan Richards; Zachary Brown; Laura Blue; Mani A Daneshmand; Jacob N Schroder; Dawn E Bowles; Carmelo A Milano
Journal:  J Vis Exp       Date:  2018-05-11       Impact factor: 1.355

2.  Driver electronics design and control for a total artificial heart linear motor.

Authors:  Kristin Unthan; Elena Cuenca-Navalon; Benedikt Pelletier; Thomas Finocchiaro; Ulrich Steinseifer
Journal:  Med Biol Eng Comput       Date:  2018-01-27       Impact factor: 2.602

Review 3.  The ongoing quest for the first total artificial heart as destination therapy.

Authors:  Annemijn Vis; Maziar Arfaee; Husain Khambati; Mark S Slaughter; Jan F Gummert; Johannes T B Overvelde; Jolanda Kluin
Journal:  Nat Rev Cardiol       Date:  2022-06-06       Impact factor: 32.419

4.  The HeartMate 6.

Authors:  Mani A Daneshmand; Muath Bishawi; Carmelo A Milano; Jacob N Schroder
Journal:  ASAIO J       Date:  2020-03       Impact factor: 3.826

5.  Design and Evaluation of a Fully Implantable Control Unit for Blood Pumps.

Authors:  Kristin Unthan; Felix Gräf; Marco Laumen; Thomas Finocchiaro; Christoph Sommer; Hermann Lanmüller; Ulrich Steinseifer
Journal:  Biomed Res Int       Date:  2015-10-25       Impact factor: 3.411

6.  Design of a Mechatronics Model of Urinary Bladder and Realization and Evaluation of Its Prototype.

Authors:  Zhao Zhuo; He Dan; Li Gensong; Shi Ping; Pang Xining
Journal:  Appl Bionics Biomech       Date:  2019-12-14       Impact factor: 1.781